The KingFisher Apex Dx instrument can only be connected in the licensed version of the Diomni Enterprise Software. Before you begin, verify that you have connected at least 1 qPCR instrument and activated the license. Refer to the system configuration for recommended Local Area Netork (LAN) setup up.
Troubleshooting steps include the following:
- Work with your IT or system administrator to set up the required shared folder in the Diomni host/server computer.
- Verify that a reserved ethernet port is connected to the KingFisher Apex Dx instrument through a LAN.
- Verify that the KingFisher Apex Dx instrument time is set to match the time displayed on the host/server computer where the SAE Administrator Console Dx software is installed (± 5 mins).
If you are still unable to connect, contact Thermo Fisher Scientific remote hardware support at instrumenthardwaresupport@thermofisher.com.