TSQ Quantis™ 三重四极杆质谱仪

以极佳的灵敏度实现稳健的目标物定量工作流程,适用于所有分子类型和不同专业级水平的用户操作

Have Questions?
货号 TSQ02-10001
价格(CNY)
-
申请报价

轻松解答实验室目标物定量工作流程中的关键分析挑战。Thermo Scientific™ TSQ Quantis™ 三重四极杆质谱仪采用分段四极杆、更快的杆驱动器和增强型双模电子倍增器,为常规分析和每个用户提供前所未有的耐用性和极佳的灵敏度。

TSQ Quantis 质谱仪能够与行业领先的定量软件、HPLC、UHPLC 以及样品制备解决方案实现无缝集成;它还为各种分子形态的目标物定量工作流程提供了更高的效率。其他优点包括:

  • 极佳的灵敏度,即使在复杂的基质中也可以获得所有分子类型的全面数据。
  • 卓越的耐用性,能够提高数据的可靠性,同时延长仪器的正常运行时间。
  • 可靠性和可重现性提高了每次运行以及每份样品的数据质量。
  • 与专业应用软件紧密集成,确保提高所有应用领域的工作效率。
  • 简单易用,帮助不同专业级水平的用户获得高质量的数据,并提高分析结果的可靠性。
规格
适用于(应用)制药、环境、食品安全、临床研究和法医毒理学
质量范围5 至 3000 m/z
类型三重四极杆 LC-MS
Unit SizeEach

常见问题解答 (FAQ)

How do I clean/store my Triple Quadrupole mass spectrometry instruments after analysis?

For short periods (less than 24 hr) the system can just be put into standby. However, it is good practice to flush the system with 50:50 solvent/water (where the solvent is either methanol or acetonitrile) to remove buffers or additives and stop microbial growth.

For longer periods, flush the LC system with 50:50 solvent/water. Then flush with 100% solvent. Put the system into standby.

Find additional tips, troubleshooting help, and resources within our Mass Spectrometry Support Center.

How do I perform mass calibration on Triple Quadrupole mass spectrometry instruments and how often should I do it?

The mass calibration should ideally be performed every 3 months and definitely performed at least every 6 months. Refer to the Instrument Manual (https://tools.thermofisher.com/content/sfs/manuals/TSQ-Series-Hardware.pdf) for detailed instructions on how to perform mass calibration.

Find additional tips, troubleshooting help, and resources within our Mass Spectrometry Support Center.

How should I clean the Ion Transfer Tube on Triple Quadrupole mass spectrometry instruments?

The Ion Transfer Tube should be removed and sonicated in a 50:50 solution of methanol/water that contains 20% formic acid for 30 min. Rinse thoroughly with water. Then sonicate in methanol for 15 min. Dry with nitrogen gas.

The same procedure can be applied to the Ion Sweep Cone.

Refer to the Instrument Manual (https://tools.thermofisher.com/content/sfs/manuals/TSQ-Series-Hardware.pdf) for more detailed instructions.

Find additional tips, troubleshooting help, and resources within our Mass Spectrometry Support Center.

How should I adjust my ion source position on Triple Quadrupole mass spectrometry instruments?

Perform loop injections of your analyte using isocractic elution through a short column at the flow rate and approximate mobile phase composition. Use a concentration where the signal to noise is approximately 10:1. Record the raw data file and integrate the peaks, comparing peak areas at different source positions (i.e., left and right, up and down, fore and aft). Ensure that the background noise is not increasing more than the peak area.

Note: The needle position on the Thermo Fisher Scientific TSQ Altis MS, TSQ Quantis MS, and TSQ Fortis MS instruments is set and should not be adjusted.

Find additional tips, troubleshooting help, and resources within our Mass Spectrometry Support Center.

Do I need to warm/equilibrate Triple Quadrupole mass spectrometry instruments prior to starting a session of analysis?

It is advisable to turn both the LC system and the MS system on for at least 30 min prior to analysis. This time can usually be spent testing one or more solvents or matrix blanks.

Find additional tips, troubleshooting help, and resources within our Mass Spectrometry Support Center.