Orion™ Epoxy Bodied Polarographic Dissolved Oxygen Electrodes
Orion™ Epoxy Bodied Polarographic Dissolved Oxygen Electrodes
Actual product may vary
Orion™ Epoxy Bodied Polarographic Dissolved Oxygen Electrodes
Orion™ Epoxy Bodied Polarographic Dissolved Oxygen Electrodes
Orion™ Epoxy Bodied Polarographic Dissolved Oxygen Electrodes
Orion™ Epoxy Bodied Polarographic Dissolved Oxygen Electrodes
Thermo Scientific™

Orion™ Epoxy Bodied Polarographic Dissolved Oxygen Electrodes

Get accurate dissolved oxygen readings, effortlessly, in the lab or on-site.
Have Questions?
Change viewbuttonViewtableView
Catalog NumberLength (English) Cable
083005MD4.9 ft.
083010MD9.8 ft.
083025MD32.81 ft.
Catalog number 083005MD
Price (CNY)
-
Length (English) Cable:
4.9 ft.
Thermo Scientific Orion polarographic DO electrode 083010MD complies with SM 4500-O G, ASTM D888-09 (B).
 
Featuring automatic temperature compensation for accurate readings, the polarographic dissolved oxygen electrodes also offer convenient screw-on replacement membrane caps. This design significantly reduces electrode servicing time and eliminates common assembly issues. The versatility of these electrodes makes them compatible with a wide spectrum of applications, including water and wastewater analysis, beer and wine testing, and demanding field work. The electrodes are available with cable lengths ranging from 1.5 to 10 meters.
 
Solutions required
Electrolyte
080514     Polarographic electrolyte solution, 60ml
 
Accessories
080513     Maintenance kit includes 2 membrane caps, electrolyte solution 080514 and polishing disk
080515     1 Membrane cap
080017     Calibration sleeve
080360     BOD adapter
081045     Plastic probe guard
 
Items included with 
080514     Polarographic electrolyte solution. 60ml 1 bottle
080515     1 Membrane cap
 
Warranty period
12 months from date of purchase
Specifications
Connections9-pin MiniDIN
DescriptionReduce probe servicing time and eliminate assembly problems
Drift∼0.1%/day
For Use With (Equipment)Orion Star Series Meters
Length (English) Cable4.9 ft.
Length (Metric) Cable1.5 m
Max. Overpressure6 bar
Min. Sample Flow10 cm/sec.
Oxygen Consumption0.008μg/h (mg/L)-1 at 20°C
Response Time90% of final value in 10 sec.
Storage Requirements-5°C to +50°C
Sensor TypesDual Thermally Separated
Temperature (Metric) Operating0°C to 50°C
TypeDO Probe, Mini-DIN Connector
Unit SizeEach