| Catalog Number | Type |
|---|---|
| 222-265600 | Cuvette Rack |
| 840-250500 | Cell Holder |
| 840-250600 | Cylindrical Cell Holder |
| 840-250700 | Replacement Sample Compartment |
| 269-268100 | UV-Vis Spectrum Poster |
| 840-253100 | SPECTRONIC Standards 2 Kit |
| 840-234000 | Thermal Printer |
Rugged, Practical Design
The Perfect Tool for Routine Spectroscopic Measurements
Not ready to part with your old Thermo Scientific Spec 20 or GENESYS 20
Check the performance and accuracy of your spectrophotometer right in your own laboratory using the Thermo Scientific™ SPECTRONIC™ Spectrophotometer Standards. This kit is as easy to use as a standard cuvette and contains:
Four calibrated neutral density glass standards for checking photometric accuracy at nominal values of 0.3A, 0.5A, 1A and 1.5A Standards are calibrated at 440.0n, 465.0nm, 546.1nm, 590.0nm and 635.0nm
One calibrated standard for checking wavelength accuracy at nominal wavelengths on 400nm, 522nm and 777nm
The symmetry of the peaks means that tests of wavelength accuracy are independent of the spectral bandwidth (between 1nm and 10nm) of the instrument being tested
1 filter for checking stray light in UV-Vis instruments at 220nm
1 filter for checking stray light in Vis instruments at 400nm
Each set of SPECTRONIC Standards is individually tested and certified to be accurate, with values traceable to the National Institute of Standards and Technology (NIST) or other Primary Standards
Filters are supplied in a tough plastic storage container
Recalibration service is available from Thermo Fisher Scientific. Recommended recalibration interval is two years
Note: Stray light filters should block light completely at the measurement wavelength and are therefore not something that can be calibrated.