The Ion OneTouch™ DL Upgrade Kit enables a one-time 'DL' upgrade to the Ion OneTouch™ System, allowing use of theRead more
Have Questions?
Catalog Number
Quantity
4474523
1 Kit
Catalog number 4474523
Price (CNY)
-
Quantity:
1 Kit
The Ion OneTouch™ DL Upgrade Kit enables a one-time 'DL' upgrade to the Ion OneTouch™ System, allowing use of the Ion OneTouch™ 200 Template Kit v2 DL. The Ion OneTouch™ DL Upgrade Kit is customer-installable, with the upgrade taking no more than a few minutes. The kit includes: injector hub, tubing catch, wrench, and firmware upgrade loaded onto a USB flash drive. Only 1 upgrade kit is required per Ion OneTouch™ System.
The Ion OneTouch™ DL Upgrade Kit provides you with:
• Improved instrument robustness • Improved instrument performance • Compatibility with the Ion OneTouch™ 200 Template Kit v2 DL Kit
Note: This upgrade must be performed for the OneTouch™ System to be compatible with the Ion OneTouch™ 200 Template Kit v2 DL.
For Research Use Only. Not for use in diagnostic procedures.
Specifications
For Use With (Equipment)Ion OneTouch™ System
Product TypeDL Upgrade Kit
Quantity1 Kit
Shipping ConditionRoom Temperature
Unit SizeEach
Contents & Storage
Includes injector hub, tubing catch, wrench, and firmware upgrade loaded USB flash drive.
Store at room temperature.
Frequently asked questions (FAQs)
What should I do if the centrifuge is still running even though the time remaining is 0:00 at the final spin of the Ion OneTouch 2?
The "time remaining" is not entirely accurate. The centrifuge may keep rotating for about 15 minutes after the time has become 0:00. If it rotates for more than 15 minutes, please contact Technical Support.
What should I do if gas leaks from the Ion OneTouch DL after I open the gas main?
Please check the following:
1) That the connection tube between Ion OneTouch DL is not loose
2) That the outlet of the gas tank regulator is at 30 psi
If the tube is not loose and the regulator value is correct, please pull the valve (the gold ring) behind the Ion OneTouch DL straight down a couple of times.
If the pressure is too high, you will need to make adjustments. For details, please contact Technical Support.
What are these black precipitates in the recovery tube and waste of the Ion OneTouch System?
It may be rubber from the packing. The amount varies among instruments and it is not anything to be concerned about, as it will be removed by the Ion OneTouch ES.