Specifications
Auxiliary GasFuel gas: Hydrogen, 30 ml/min, 0.2MPa, Purity > 99.999%, THC< 0.1 ppm
Combustion-supporting gas: Air, pure, oil-free and water-free, 300 ml/min, 0.2MPa,THC< 0.1 ppm
Carrier gas: Nitrogen, pure, oil-free and water-free, 60 ml/min, 0.2MPa, THC<0.1 ppm
Zero calibration gas: Nitrogen, pure, oil-free and water-free, 0.2MPa, THC< 0.1 ppm, (carrier gas is nitrogen)
Range calibration gas: Hydrocarbons of known concentration equilibrated in nitrogen (nitrogen as carrier gas), inflow pressure is 0.1MPa
Pipe purge gas: Air, pure, oil-free and water-free, 10 L/min, 0.2MPa, THC< 1 ppm
Calibrationdaily - weekly, can be set by users
Communication Ports1 x RS-232, 1 x USB, 2 x network ports (optional: 4-20 mA)
DescriptionVOC6000 Analyzer (BTEX)
Electrical RequirementsAnalysis system: AC 220V, 50/60 Hz, 2 kW (Including system cabinet +sampling probe)
Sampling pipe: AC 220V, 50/60 Hz, 3.25 kW (heat tracing pipe of 50m length)
Measurement Range0-50ppm/500ppm/5000ppm/5%/50% in methane
OutputsInstrument calibration
verification/instrument
calibration/instrument
maintenance/instrument
abnormality/hydrogen cutoff
Sampling Flow and Control100-1000 mL/min
Measurement MethodGC-FID, Hydrogen Flame Ionization Method
Temperature (English) Operating32°F to 113°F, ≤85% RH (no condensation); Sample temperature: 59°F to 428°F, 85% RH (no condensation)
Temperature (Metric) Operating0°C to 45°C, ≤85% RH (no condensation), Sample temperature: 15°C to 220°C, 85% RH (no condensation)
Unit SizeEach