Attune™ NxT Maintenance Kit
Attune™ NxT Maintenance Kit
Invitrogen™

Attune™ NxT Maintenance Kit

The Attune NxT Maintenance Kit will help you avoid down-time and ensure optimal operation of your Attune NxT Acoustic FocusingRead more
Have Questions?
Catalog NumberQuantity
A430381 kit
Catalog number A43038
Price (CNY)
-
Quantity:
1 kit
The Attune NxT Maintenance Kit will help you avoid down-time and ensure optimal operation of your Attune NxT Acoustic Focusing Cytometer hardware. This convenient kit includes sufficient supplies to perform all recommended preventative maintenance procedures for 12 months. These procedures are: replacement of the syringe (bi-yearly), SIP tube (yearly), and instrument bottles (yearly), as well as performance of system decontamination (quarterly). The kit includes 1 waste bottle, 1 focusing fluid tank, 12 focusing fluid filters, 2 syringes, 1 SIP tube, and 1 debubble solution.
For Research Use Only. Not for use in diagnostic procedures.
Specifications
Block FormatNon-Interchangeable
Product LineAttune
Quantity1 kit
Shipping ConditionApproved for shipment at Room Temperature or on Wet Ice
For Use With (Equipment)Attune Acoustic Focusing Cytometer
Product TypeMaintenance Kit
Unit SizeEach
Contents & Storage
• 1 waste bottle
• 1 focusing fluid bottle
• 12 focusing fluid filters
• 2 syringes
• 1 SIP
• 1 debubble solution

Store at room temperature.

Frequently asked questions (FAQs)

How often should I change filters for the Attune NxT Acoustic Focusing Cytometer?

Filters are recommended for replacement once every three months, unless performing small particle detection when once-a-month replacement is recommended.

Find additional tips, troubleshooting help, and resources within our Cell Analysis Support Center.

Can I purchase the individual components of the Attune NxT Maintenance Kit as standalone products?

Yes. Here are the catalog numbers for the individual components:
- Cat. No. 100022156: 1 Waste bottle
- Cat. No. 100022155: 1 Focusing fluid bottle
- Cat. No. 100022587: 12 Focusing fluid filters
- Cat. No. 100022591: 2 Syringes
- Cat. No. 4490099: 1 SIP
- Cat. No. A10496: Debubble solution

Find additional tips, troubleshooting help, and resources within our Cell Analysis Support Center.

Is the debubble solution provided in the Attune NxT Maintenance Kit the same as the standalone product, Attune Debubble Solution (1X), Cat. No. A10496?

Yes. It is the same 50 mL bottle of 1X Debubble Solution as provided in Attune Debubble Solution (1X).

Find additional tips, troubleshooting help, and resources within our Cell Analysis Support Center.