DFS™ Magnetic Sector GC-HRMS System
DFS™ Magnetic Sector GC-HRMS System
DFS™ Magnetic Sector GC-HRMS System
DFS™ Magnetic Sector GC-HRMS System
Thermo Scientific™

DFS™ Magnetic Sector GC-HRMS System

Achieve outstanding sensitivity for dioxin and persistent organic pollutants (POPs) analysis using this gold-standard GC-HRMS system.

Have Questions?
Catalog NumberSensitivity
IQLAAEGAALFABKMAEGSN ≥200:1 (PtP, 4σ) for 20 fg 2,3,7,8-Tetrachlorodibenzo-p-Dioxin on Column
Catalog number IQLAAEGAALFABKMAEG
Price (CNY)
-
Request A Quote
Sensitivity:
SN ≥200:1 (PtP, 4σ) for 20 fg 2,3,7,8-Tetrachlorodibenzo-p-Dioxin on Column

Meet your current and future needs for worldwide compliance in Dioxins and POPs analysis with the Thermo Scientific™ DFS Magnetic Sector GC-HRMS, which delivers ultimate confidence and robustness. Software features support fully compliant, routine quantification of POPs, with the automation required to meet your lab needs.

DFS Magnetic Sector GC-HRMS System

  • Based on magnetic sector technology, the DFS Magnetic Sector GC-HRMS performs dioxin and POPs routine analysis, delivering ultimate confidence, no matter the sample type or matrix.
  • Offers worldwide full compliance with any official dioxin, PCB, or PBDE method (e.g., EPA 1613, 1668, 1614).

Thermo Scientific DFS Software Package

  • Includes automated features supporting fully compliant routine quantifications of POPs, with the automation required to meet your lab needs. Designed with ease-of-use in mind, the Thermo Scientific™ DFS™ Software Package ensures full data confidence while delivering maximum flexibility, even in regulated labs.
  • Thermo Scientific™ DFS SmartTune™ Operating Software streamlines everyday tasks, from instrument optimization and troubleshooting to routine method development.
  • Thermo Scientific™ DFS TargetQuan™ Data Evaluation Software facilitates the quantification of dioxins and other POPs in accordance with isotope dilution requirements.
Specifications
Data System Software and OptionsThermo Scientific Xcalibur and TargetQuan 3
DescriptionDFS High Resolution GC/MS, Configurable
Dynamic Range>5 Orders Linear Dynamic Range
Mass Accuracy<2 ppm Voltage Scanning, <5 ppm Magnet Scanning
ModesFull Range of Ionization Modes Available: EI, CI, and FD/FI
Scan Rates0.1 to 10,000 Da/sec.
Scanning CapabilitiesVoltage Scanning, Magnet Scanning, Mulitiple Ion Detection (MID Aka SIM), Multiple Reaction Monitoring (MRM)
SensitivitySN ≥200:1 (PtP, 4σ) for 20 fg 2,3,7,8-Tetrachlorodibenzo-p-Dioxin on Column
Source TypeOptimized Ion Efficiency, Easy Install, Inert Source
Vacuum SystemTurbo Molecular Pumps Backed by Rotary Pumps, All Housed Inside The Instrument for Minimum Noise Emission
Voltage230 V
Mass Range2 to 6000 m/z
Resolution>60,000 R
TypeGC-HRMS System
Unit SizeEach