Specifications
Analysis Time0.3 to 1.0 sec. per gas component
Analyzer TypeScanning Magnetic Sector Mass Spectrometer with Enclosed Electron Impact Dual Filament Ion Source
Area Classification OptionsGeneral Purpose; Z-purged Class 1 Div 2 (optional); ATEX Zone 1, IIC T3 (optional)
Depth (English)2.7 ft.
Depth (Metric)0.8 m
Dynamic RangeFaraday: 10 ppm to 100 ppm
Optional single MCP SEM: 1 ppm to 1000 ppm
Optional twin MCP SEM: 10 ppb to 1000 ppm
Height (English)GP: 5 ft., Ex: 6 ft.
Height (Metric)GP: 1.5 m, Ex: 1.8 m
No. of Calibration Gas InletsUp to 6 (standard); optional to 24
No. of Sample Stream InletsUp to 31 or 63 (standard); optional to 126
Power115/230 V, Consumption GP: 2.0 kVA; Ex: 2.5 kVA
Precision≤ 0.1% relative (typical, application dependent)
Stability≤ 1% relative over one month (typical, application dependent)
Weight (English)530 lb.
Weight (Metric)240 kg
Width (English)3.3 ft.
Width (Metric)1 m
Ambient Temperature12°C to 40°C
Communication ProtocolsModbus serial RS232/422/485, Siemens 3964R, VGCP and PVGCP; optional via protocol converters: Modbus Ethernet TCP/IP and Profibus DP
Mass RangeAdjustable, 1 to 150 amu at 1000 eV ion acceleration voltage, 1 to 200 amu at 750 eV ion acceleration voltage
Product TypePrima PRO Process Gas Analysis Mass Spectrometer
Unit SizeEach