DS-30 基质标准品试剂盒(染料组 D)
DS-30 基质标准品试剂盒(染料组 D)
Applied Biosystems™

DS-30 基质标准品试剂盒(染料组 D)

DS-30 基质标准品试剂盒包含四种分别带有 6-FAM、HEX、NED 和 ROX 标记的寡核苷酸,并在单个管中预混合。本款基质标准品套装用于生成在 Applied了解更多信息
Have Questions?
货号数量
43458278 runs
货号 4345827
价格(CNY)
2,992.00
Each
添加至购物车
数量:
8 runs
价格(CNY)
2,992.00
Each
添加至购物车
DS-30 基质标准品试剂盒包含四种分别带有 6-FAM、HEX、NED 和 ROX 标记的寡核苷酸,并在单个管中预混合。本款基质标准品套装用于生成在 Applied Biosystems 系统上分析荧光染料标记的 DNA 片段时所需的“多组分基质”。用于这些仪器的数据采集软件使用多组分基质在单个毛细管中自动分析经四种不同颜色的荧光染料标记的样品。
仅供科研使用。不可用于诊断程序。
规格
适用于(设备)SeqStudio 3500 and 3730xl
数量8 runs
建议的储存条件Contains 4 oligonucleotides labeled respectively with 6FAM™, HEX™, NED™, and ROX™. For Dye Set D spectral calibration. Store the kit at 2 - 8°C. Do not freeze.
运输条件Wet Ice
Unit SizeEach

常见问题解答 (FAQ)

What is a spectral calibration?

A spectral calibration is an algorithm applied to raw data, which converts it into the component 4 or 5 dye data stored in the sample files. A spectral is created for a specific dye set (combination of dyes), array type (4 or 16 capillaries), and array length (36cm or 50cm). It is used to correct for the natural overlap of the fluorescent dyes.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Instruments Support Center.

Which fragment analysis matrix standards can I use for my Applied Biosystems 3130 Series instrument?

DS-02 (Dye Set E5), DS-30 (Dye Set D), DS-31 (Dye Set D), DS-32 (Dye Set F), and DS-33 (Dye Set G5) are all supported on the Applied Biosystems 3130 Series systems. Please refer to the Applied Biosystems 3130/3130xl Genetic Analyzers Getting Started Guide (http://tools.thermofisher.com/content/sfs/manuals/cms_041468.pdf) for more information.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Applications Support Center.

I am using only one dye for fragment analysis but I see peaks in other colors below my peak of interest. Why is this?

If the peaks in other colors are directly below the peak of interest, the issue could be that the fluorescent dye being used is not part of the selected dye set, the spectral calibration needs to be performed, or the peaks are offscale. Confirm that the dye set selected on the instrument is compatible with the dye being used, run a new spectral calibration if the correct dye set has been selected and, if the signal intensity is too high, decrease sample concentration during PCR or when preparing samples for electrophoresis.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Applications Support Center.

Can I switch dyes during a fragment analysis project?

We do not recommend that you switch dyes in the middle of a project. Switching dyes may result in a shift in the allele sizes as dyes differ in their mobility. If dyes are switched in the middle of a project, it will be necessary to identify the size shift for all the markers/targets in the amplicon product range. The size shift will be consistent throughout the project and it is only necessary to make the adjustment once.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Applications Support Center.

What is a dye set?

A dye set is a combination of dyes that have minimum spectral overlap, which will allow you to multiplex products of similar size using different dyes. Although the size ranges may overlap, the use of a different dye will allow the software to easily distinguish between these products.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Applications Support Center.