测序标准品,BigDye™ Terminator v1.1,适用于 3500/3500xL Dx
测序标准品,BigDye™ Terminator v1.1,适用于 3500/3500xL Dx
Applied Biosystems™

测序标准品,BigDye™ Terminator v1.1,适用于 3500/3500xL Dx

该测序标准品包含4管,每管均含有使用 BigDye Terminator v1.1 制备的1,200个已知碱基对序列的 DNA。所含的 DNA 足以在 Applied了解更多信息
Have Questions?
货号数量
44621134 管
货号 4462113
价格(CNY)
-
数量:
4 管
该测序标准品包含4管,每管均含有使用 BigDye Terminator v1.1 制备的1,200个已知碱基对序列的 DNA。所含的 DNA 足以在 Applied Biosystems 3500 Dx 系列基因分析仪上执行一次光谱校准/性能检查和/或一次对照测序运行。DNA 制备品已经过冻干处理以尽可能提高稳定性。
仅供研究使用
规格
适用于(应用)测序
适用于(设备)3500xL Dx 基因分析仪、3500 Dx 基因分析仪、3500xL 基因分析仪、3500 基因分析仪
标签或染料ROX,TAMRA™,罗丹明 110,罗丹明 6G
产品线BigDye™ Terminator
产品类型3500 Dx 系列测序标准品
数量4 管
Unit SizeEach
内容与储存
该试剂盒含有 4 管使用 BigDye™ Terminator v1.1 制备的 1200 个已知碱基对序列的 DNA。每管可在 Applied Biosystems 3500 Dx 系列基因分析仪上执行一次光谱校准/性能检查和/或一次对照测序运行。将试剂盒储存在 -15 至 -25°C 下。

常见问题解答 (FAQ)

When I started my run on the Applied Biosystems 3500/3500xL Genetic Analyzer, an error occurred: "Unstable Electrophoresis current detected, check for air bubbles" but I don't see any bubbles. What else can cause this?

Some of the other causes of an Unstable Electrophoresis current detected error message are:

1.Leak on the system
2. Polymer that:
-Has expired
-Has been left on the instrument for more than the recommended time
-Is a mixture of expired polymer and non-expired polymer
3. Running Buffer that was:
-On the instrument longer than 2 weeks
-Not filled to the fill line or evaporated below the fill line
4. An arcing event that was not cleaned afterwards using the water wash wizard
5. Not performing regular maintenance on the instrument
6. Hardware issues

Inspect the system for leaks. If you do not see any leaks on the system, perform the wash the pump chamber and channels wizard using the conditioning pouch and place fresh, non-expired polymer and fresh Anode and Cathode buffer containers. If the problem persists, a service call may be required.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Instruments Support Center.

Do I need to replace the buffer every 14 days if the Applied Biosystems 3500/3500xL Genetic Analyzer is not in use?

If the instrument is not in use, it is not necessary to replace the buffer every 14 days. However, the buffer level needs to remain at the fill line to prevent the capillary or array from drying out. Top off the buffer volume with water if necessary. Replace the anode and cathode buffer containers prior to starting a new run.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Instruments Support Center.