Specifications
Diameter (English) WindowApproximately 0.45 in. with frame
Diameter (Metric) WindowApproximately 11.4 mm with frame
Humidity RangeAmbient Temperature and Humidity: 0 to 35°C, 0-95% RH, up to 5,000 feet altitude
Input Power28-31 VDC, 3 amps
Max. Beam CurrentDXS11-5025: 0.50 mA, 43-50kV
DXS11-8012: 0.10 mA at 20kV, 0.15 mA, 60-80kV
Max. PowerDXS11-5025: 25W at 50kV
DXS11-8012: 1 2W at 80kV
Method of CoolingInternal fan is sufficient for ambient temperature up to 28°C ambient and up to 12W operation. Above 28°C ambient, or for continuous operation above 12W, cooling air must be directed at the unit.
Operating Voltage RangeDXS11-5025: 10-50kV
DXS11-8012: 40-80kV
Shielding TypeShielding is limited. User must provide adequate shielding to ensure safe operation.
Spot Elipticity< 20% typical
Spot SizeDXS11-5025: 65μ at 50kV 25W
DXS11-8012: 30μ at 80kV 12W
Spot to Window Spacing8.9 mm (0.35 in.)
TypeDXS11 High Resolution Mini-focus X-Ray Source
Weight (English)Approximately 11.5 lb.
Weight (Metric)Approximately 5.2 kg
Cone of Illumination34° x 49°
Power Consumption3 amps
Target MaterialTungsten
Window MaterialBeryllium
Unit SizeEach