Thermo Scientific PXS11 -75kV Microfocus X-Ray Source
Thermo Scientific PXS11 -75kV Microfocus X-Ray Source
Thermo Scientific™

Thermo Scientific PXS11 -75kV Microfocus X-Ray Source

Thermo Scientific™ PXS11 - 75kV Microfocus X-Ray Source is an affordable solution for intermittent use in industrial, healthcare and laboratory applications, combining a stable, low energy output and moderate resolution.
Have Questions?
Catalog NumberTube Current Operational Range
PXS110 – 100 µA
Catalog number PXS11
Price (CNY)
-
Request A Quote
Tube Current Operational Range:
0 – 100 µA

Thermo Scientific PXS11 – 75kV Microfocus X-Ray Source is an affordable solution for intermittent use in industrial, healthcare and laboratory applications, combining a stable, low energy output and moderate resolution. With a round focal spot for accurate imaging at the cell cluster scale it is particularly suitable for medical fluoroscopy and techniques such as tomosynthesis; alternative applications include component inspection and non-destructive testing. The PXS11 X-Ray Source is Restriction of Hazardous Substances (RoHS) compliant with a simple, durable, compact design that is easy to install and highly reliable.

PXS11 75kV Microfocus X-Ray Source is specified for low-cost applications requiring moderate to high resolution. Key features include:

  • Simple, durable design, for easy installation, high reliability, and longevity
  • Auto-conditioning, for protection of the source during start-up
  • Round spot, for high quality, low distortion imaging
  • Fully integrated design with x-ray tube, power supply, and controller in a single compact package, for easy system integration
  • Low energy output, for enhanced safety in medical applications
  • Side window x-ray tube, for installation in compact systems
Specifications
Ambient Humidity0 - 95% RH, up to 1,500 m (5,000 ft) altitude
Ambient Temperature0 - 30 °C
Control InterfaceAnalog control and monitoring of operating conditions and status
Focus to Object Distance (FOD)8.9 mm
Input Power28 - 30 VDC, 3A max
Maximum Power Output7.5 W
Method of CoolingInternal fan. Cooling air should be provided for continuous operation.
Minimum Focal Spot Size45 µm
Operating Voltage Range20 -75 kV
Tube Current Operational Range0 – 100 µA
Tube Voltage75 kV Max.
Weight (Metric)4.0 kg
X-ray Beam Angle34o, round beam
X-ray Output Window MaterialBeryllium
Target MaterialTungsten
Unit SizeEach