DynaMag™-96 Bottom Magnet - FAQs

View additional product information for DynaMag™-96 Bottom Magnet - FAQs (12332D)

12 product FAQs found

DynaMag-2和DynaMag-Spin磁力架的高斯值是多少?

DynaMag-2磁力架的最低高斯值为3,500–3,700,DynaMag-Spin磁力架最低高斯值为3,000–3,500。

高斯和特斯拉单位的标准换算关系是多少?

1特斯拉等于10,000高斯。

我该如何清洁DynaMag磁力架?

使用以下任意一种清洁剂喷洗和/或擦拭DynaMag磁力架:

•70%异丙醇
•1%次氯酸钠溶液(漂白剂)
•0.1 N HCl溶液

不要将装置浸泡在水溶液中,避免长时间接触于水或其他水溶液。不可将磁力架进行高压灭菌或加热至70°C以上。

DynaMag-2磁力架(货号12321D)和DynaMag-Spin磁力架(货号12320D)有何区别?

DynaMag-2磁力架可容纳16个标准的1.5 mL或2 mL微量离心管,而DynaMag-Spin磁力架最多能容纳6个标准的1.5 mL微量离心管。

What are the Gauss levels for the DynaMag-2 and DynaMag-Spin Magnets?

The DynaMag-2 Magnet is 3,500-3,700 minimum Gauss level and the DynaMag-Spin is 3,000-3,500 minimum Gauss level.

Find additional tips, troubleshooting help, and resources within our Protein Assays and Analysis Support Center.

What is the standard conversion of Gauss to Tesla units?

1 Tesla is equivalent to 10,000 Gauss. .

How can I clean my DynaMag magnet?

Spray and/or wipe the DynaMag magnet unit with one of the following cleaning agents:

- 70% (v/v) isopropyl alcohol
- 1% (v/v) sodium hypochlorite solution (bleach)
- 0.1 N HCl solution

Do not submerge the magnet unit in aqueous solutions and avoid prolonged exposure to water or aqueous solutions. Clean with a damp cloth and mild detergent when exposed to harsh solvents. Do not autoclave the magnet unit as heating the magnet to above 70 degrees C can destabilize the magnetic material, resulting in loss of magnetic strength.

What is the difference between the DynaMag-2 Magnet (Cat. No. 12321D) and DynaMag-Spin Magnet (Cat. No. 12320D)?

The DynaMag-2 Magnet can hold 16 standard 1.5 mL or 2 mL microcentrifuge tubes, while the DynaMag-Spin Magnet holds up to 6 standard 1.5 mL microcentrifuge tubes.

What is the definition of superparamagnetic, and what does this mean for my cell isolation application with Dynabeads magnetic beads?

Superparamagnetic means that the Dynabeads magnetic beads exhibit magnetic properties when placed within a magnetic field, but have no residual magnetism when removed from the magnetic field.

This means that your targeted cells, proteins, or nucleic acids are only subjected to magnetic forces during the time the beads are on the magnet. The beads do not aggregate, but remain evenly dispersed in suspension.

Find additional tips, troubleshooting help, and resources within our Dynabeads Cell Isolation and Expansion Support Center.

Are the antibodies on your Dynabeads magnetic beads for cell isolation/activation/expansion covalently bound to the beads?

Yes. The antibodies are covalently bound and should be very stable in your applications.

Find additional tips, troubleshooting help, and resources within our Protein Assays and Analysis Support Center.

What is the average shelf life of Dynabeads magnetic beads?

Depending on the antibody coated on the Dynabeads magnetic beads, the shelf life can vary from 24-36 months.
Some kits may have 18 months shelf life depending on other components supplied in the kit. The kits are guaranteed for 6 months from when you receive them.

Find additional tips, troubleshooting help, and resources within our Dynabeads Nucleic Acid Purification Support Center as well as our Protein Immunoprecipitation (IP), Co-Immunoprecipitation (Co-IP), and Pulldown Support Center.

How can I clean my DynaMag magnet?

Spray and/or wipe the DynaMag magnet unit with one of the following cleaning agents:

- 70% (v/v) isopropyl alcohol
- 1% (v/v) sodium hypochlorite solution (bleach)
- 0.1 N HCl solution

Do not submerge the magnet unit in aqueous solutions and avoid prolonged exposure to water or aqueous solutions. Clean with a damp cloth and mild detergent when exposed to harsh solvents. Do not autoclave the magnet unit as heating the magnet to above 70 degrees C can destabilize the magnetic material, resulting in loss of magnetic strength.

Find additional tips, troubleshooting help, and resources within our Protein Assays and Analysis Support Center.