CalibrationSource test routine checks alpha/beta efficiency and crosstalk values. Efficiencies automatically computed based upon user-defined sources. Automatic decay correction of sources. Source tests are automatically prompted at user-defined intervals.
Certifications/ComplianceIP 32 according to EN 60 529
Check SourceRoutine that permits quick verification and operability of system to user-defined acceptance criteria. All check source events are logged.
Count RateUser-selectable between 1 and 9999 seconds
Crosstalk< 5% alpha into beta < 1% beta into alpha
DescriptionRadEye HEC+ Alpha/Beta Sample Counter
Detector Type50.8 mm diameter (20 cm²), windowless alpha and beta-sensitive scintillator
Dimensions (L x W x H)120 x 380 x 304 mm
Humidity10% to 95% at 35°C Non-Condensing
Internal MemoryThe last 4500 measured values are saved and can be read out via PC program. Logbook with 250 entries for changes of configuration and errors.
MechanicalSingle-package design with carrying handle facilitates easy portability.
Operating Temperature (Metric)−20°C to +50°C
Power Supply100–240 VAC, 47–63 Hz Charging indicators for internal NiMH battery. Rechargeable battery, 7.2 volts, 3.9 Amp Hr., 800 hours of operation between charges.
Preset CountUser-selectable between 1 and 9999 seconds.
Sample DrawerThe sample holder and slide have a durable finish for ease of decontamination. Adapters for a variety ofcircular and rectangular air filters and test sources are provided. Maximum diameter for circular filters: 60 mm.
Unitscps, cpm, Bq, dps, dpm, Bq/cm2
Weight (English)9.0 lb.
Background MeasurementUser-selectable between 1 and 9999 seconds; used in subtraction of sample counts.
Weight (Metric)4.1 kg
Unit SizeEach