3730 DNA 分析仪 48-毛细管阵列,36 cm
Product Image
Applied Biosystems™

3730 DNA 分析仪 48-毛细管阵列,36 cm

这是一种适用于 Applied Biosystems™ 3730 DNA 分析仪的毛细管阵列。每个阵列均包含 48 个毛细管,长度为 36了解更多信息
Have Questions?
货号数量
43312471 个阵列
货号 4331247
价格(CNY)
-
数量:
1 个阵列
这是一种适用于 Applied Biosystems™ 3730 DNA 分析仪的毛细管阵列。每个阵列均包含 48 个毛细管,长度为 36 cm。用于测序以及片段分析运行。

每个阵列的用量:300 次运行(使用 Applied Biosystems 试剂的情况下)。
仅供科研使用。不可用于诊断程序。
规格
毛细管长度36 cm
适用于(应用)测序
适用于(设备)3730 DNA Analyzer
毛细管数48支毛细管
产品类型3730 DNA 分析仪 48 毛细管阵列
数量1 个阵列
运输条件室温
产品规格毛细管阵列
Unit SizeEach
内容与储存
每个阵列均包含 48 个毛细管(长度为 36 cm),与 Applied Biosystems 3730 DNA 分析仪配合使用。

我们保证在购买 1 年内最多 300 次运行中 3730 毛细管阵列的性能。

常见问题解答 (FAQ)

What is a spectral calibration?

A spectral calibration is an algorithm applied to raw data, which converts it into the component 4 or 5 dye data stored in the sample files. A spectral is created for a specific dye set (combination of dyes), array type (4 or 16 capillaries), and array length (36cm or 50cm). It is used to correct for the natural overlap of the fluorescent dyes.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Instruments Support Center.

What is the lifespan of an Applied Biosystems 3730 Series DNA Analyzer capillary array?

The life span of an array can vary depending on the application, sample prep (clean-up), and excessive handling. Each Applied Biosystems 3730 Series DNA Analyzer capillary array (regardless of length or number of capillaries) has a minimum specification of 300 runs within 1 year of purchase under normal run conditions. This minimum specification assumes careful maintenance and proper usage of the system.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Instruments Support Center.

What are the sequencing performance specifications for the Applied Biosystems 3730 Series DNA Analyzer?

On the 36 cm array:
- The Standard Sequencing run module has a run time of 60 minutes and an average read of 700 bases with a QV >20.
- The Rapid Sequencing run module has a run time of 35 minutes and an average read length of 550 bases with a QV >20.

On the 50 cm array:
- The Extra Long Read Sequencing run module has a run time of 180 minutes and an average read length of 900 bases with a QV >20.
- The Long Read Sequencing run module has a run time of 120 minutes and an average read length of 850 bases with a QV >20.
- The Fast Sequencing run module has a run time of 60 minutes and an average read length of 700 bases with a QV >20.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Instruments Support Center.

With my Applied Biosystems 3730/3730xl DNA Analyzer, I seem to be getting bubbles from the polymer bottle but the polymer level is normal. What is causing it?

It is possible that the polymer tube fitting may not be screwed into the pump properly, allowing air to get pulled in as the piston retracts. It is also possible that some debris may be stuck in the check valve, locking it in the open position. It may be necessary to detach and carefully attach the polymer tube again and then perform a water wash wizard to remove the bubbles and clear any debris from the check valve.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Instruments Support Center.

When I started my run on the Applied Biosystems 3730/3730xl DNA Analyzer, an error occurred: "Leak detected" but I don't see any leaks. What else might cause this?

“Leak detected” error messages can also occur when the piston on the pump moves further than expected during the movement of the pump. The usual cause for this is a leak on the system, usually due to either an improperly tightened or improperly threaded fitting or a broken capillary. However, if this is not observed, some of the other reasons are:

1.Valve stuck in open position
2.Large air bubble under the piston
3. Polymer:
-Wrong polymer type (e.g., instrument thinks it has POP-7 on it but it really has POP-4).
-Old/degraded polymer
4. Improper/no maintenance on the system

Inspect the system for leaks. If you do not see any leaks on the system, perform a water wash using the water wash wizard and place fresh, non-expired polymer and freshly made 1X Running Buffer on the system. If the problem persists, a service call may be required.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Instruments Support Center.