BigDye™ Terminator v1.1 和 v3.1 5X 测序缓冲液&
Product Image
Applied Biosystems™

BigDye™ Terminator v1.1 和 v3.1 5X 测序缓冲液&

此测序缓冲液经过优化,可用于 BigDye™ Terminator v1.1 和 v3.1 循环测序试剂盒。仅供研究使用。不适用于诊断用途了解更多信息
Have Questions?
更改视图buttonViewtableView
货号数量
43366971 mL
433669928 mL
4336701233 mL
货号 4336697
价格(CNY)
1,217.00
Each
添加至购物车
数量:
1 mL
价格(CNY)
1,217.00
Each
添加至购物车
此测序缓冲液经过优化,可用于 BigDye™ Terminator v1.1 和 v3.1 循环测序试剂盒。

仅供研究使用。不适用于诊断用途。
仅供科研使用。不可用于诊断程序。
规格
缓冲液测序缓冲液
最大浓度5X
适用于(应用)标准测序
适用于(设备)GeneAmp 9700、Veriti 热循环仪、3500 基因分析仪、3730xl DNA 分析仪、3130xl 基因分析仪、3500xL 基因分析仪、3730 DNA 分析仪
形式液体
产品线BigDye™
产品类型测序缓冲液
数量1 mL
运输条件室温
Unit SizeEach
内容与储存
存储在 2°C 到 8°C。

常见问题解答 (FAQ)

I am getting shorter reads than I used to. What can cause this?

A change in the read length is usually caused by something that either changes the migration of the DNA or the efficiency of the sequencing reaction. Some of the things that can cause this are:

• Buffer: Buffer may be prepared incorrectly or on the system for too long.

• Polymer: Polymer may have been on the system the system too long, used past the expiration date, frozen/crystallized and thawed, or mixed with water or buffer during capillary fill. There may be a mismatch between polymer type selected in software and what is installed on the system.

• Environment: Instrument may be operating outside environmental specifications in the Site Preparation Guide. You may have poor/improper instrument ventilation, or air flow may be blowing directly on the system.

• DNA quality: RNA/protein contamination may be reducing the sequencing reaction efficiency. If working with PCR product, there may be carryover of primers and/or dNTPs from the PCR reaction, or the sequencing reaction cleanup is incomplete. If using a kit that has beads in it for the purification, beads may be blocking the capillary tip. If using BigDye XTerminator Purification, heating the XTerminator reagent can cause loss of smaller products.

• DNA quantity: There may be too much DNA competing for entry into the capillary with labeled product, or excessive amounts of DNA creating a temporary blockage of the capillary.

For more information on other causes of short reads and how to address each issue, please refer to the DNA Sequencing by Capillary Electrophoresis: Applied Biosystems Chemistry Guide: Second Edition (Cat. No. 4305080, Rev. C). The guide can be found by searching the Thermo Fisher Scientific website with the catalog number 4305080.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Applications Support Center.

Can I dilute my BigDye Terminator reactions?

The BigDye Terminator v1.1 and v3.1 chemistries can be diluted, and a 5X Sequencing Buffer is provided with the kit to enable you to do so. However, dilution of the BigDye Terminator may compromise your data integrity depending on the cycle sequencing protocol you are following and the sequence of the particular DNA you are investigating. TheBigDye Direct Cycle Sequencing Kit and the dGTP BigDye Terminator Ready Reaction Kits v1.0 and v3.0 cannot be diluted.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Applications Support Center.

I received BigDye Terminator v1.1 & v3.1 5X Sequencing Buffer (Cat. No. 4336697) at room temperature. Is this correct?

Yes, BigDye Terminator v1.1 & v3.1 5X Sequencing Buffer is shipped at room temperature and needs to be stored at 2-8 degrees C upon receipt.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Applications Support Center.

How do I calculate the amount of BigDye Terminator v1.1 & v3.1 5X Buffer to use in the reaction?

The calculation can be done manually or by using our Applied Biosystems BigDye Terminator v1.1 & v3.1 5X Sequencing Buffer Calculator.

The formula is provided below:

Volume of BigDye Terminator v1.1 & v3.1 5X Buffer (µL) = 0.5 x [(total reaction volume/2.5) - volume of BigDye Terminator Ready Reaction Mix].

For example, if you are using 2 µL of the BigDye Terminator Ready Reaction Mix and a 15 µL reaction volume

Volume of BigDye Terminator v1.1 & v3.1 5X Buffer (µL) = 0.5 x [(15/2.5) - 2] = 2 µL

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Applications Support Center.

Can I dilute my BigDye Terminator reaction?

We provide the BigDye Terminator v1.1 & v3.1 5X Buffer to enable dilution of the BigDye Terminator Ready Reaction Mix. However, without optimization, diluting the BigDye Terminator Ready Reaction Mix may cause deterioration of sequencing quality. We cannot guarantee the performance of BigDye chemistry when it is diluted.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Applications Support Center.