基质标准品套件 DS-32 用于生成在 Applied Biosystems 3500 和 3730 系列系统上分析 5-FAM、JOE、NED 和 ROX 标记 DNA 片段时需要的“多组分基质”。数据采集软件使用多组分基质自动分析单个毛细管中的四种不同颜色的荧光染料标记样品。

支持4染料人类身份鉴定应用(即 AmpFLSTR 试剂盒)、StockMarks 和植物⁄微生物 AFLP 测定
仅供科研使用。不可用于诊断程序。
规格
适用于(应用)测序
适用于(设备)SeqStudio 基因分析仪、3500 基因分析仪、3500xL 基因分析仪、3130 基因分析仪
标签或染料FAM, JOE, NED, ROX
产品类型DS-32 基质标准试剂盒
数量8 次运行
运输条件湿冰
产品规格试剂盒
Unit SizeEach
内容与储存
包含一组具有独特荧光染料标记的基质标准 DNA 片段。足够用于 8 次运行。将试剂盒储存于 2 - 8°C 下。切勿冷冻。

常见问题解答 (FAQ)

What is a spectral calibration?

A spectral calibration is an algorithm applied to raw data, which converts it into the component 4 or 5 dye data stored in the sample files. A spectral is created for a specific dye set (combination of dyes), array type (4 or 16 capillaries), and array length (36cm or 50cm). It is used to correct for the natural overlap of the fluorescent dyes.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Instruments Support Center.

Which fragment analysis matrix standards can I use for my Applied Biosystems 3130 Series instrument?

DS-02 (Dye Set E5), DS-30 (Dye Set D), DS-31 (Dye Set D), DS-32 (Dye Set F), and DS-33 (Dye Set G5) are all supported on the Applied Biosystems 3130 Series systems. Please refer to the Applied Biosystems 3130/3130xl Genetic Analyzers Getting Started Guide (http://tools.thermofisher.com/content/sfs/manuals/cms_041468.pdf) for more information.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Applications Support Center.

I am using only one dye for fragment analysis but I see peaks in other colors below my peak of interest. Why is this?

If the peaks in other colors are directly below the peak of interest, the issue could be that the fluorescent dye being used is not part of the selected dye set, the spectral calibration needs to be performed, or the peaks are offscale. Confirm that the dye set selected on the instrument is compatible with the dye being used, run a new spectral calibration if the correct dye set has been selected and, if the signal intensity is too high, decrease sample concentration during PCR or when preparing samples for electrophoresis.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Applications Support Center.

Can I switch dyes during a fragment analysis project?

We do not recommend that you switch dyes in the middle of a project. Switching dyes may result in a shift in the allele sizes as dyes differ in their mobility. If dyes are switched in the middle of a project, it will be necessary to identify the size shift for all the markers/targets in the amplicon product range. The size shift will be consistent throughout the project and it is only necessary to make the adjustment once.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Applications Support Center.

What is a dye set?

A dye set is a combination of dyes that have minimum spectral overlap, which will allow you to multiplex products of similar size using different dyes. Although the size ranges may overlap, the use of a different dye will allow the software to easily distinguish between these products.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Applications Support Center.