调节试剂,用于 3500/SeqStudio™ Flex
Product Image
Applied Biosystems™

调节试剂,用于 3500/SeqStudio™ Flex

此处提供的调节试剂为与 3500 和 SeqStudio Flex 系列基因分析仪兼容的即用型袋。其设计用于聚合物泵的灌注,在聚合物型改变之间以及仪器关闭过程中洗涤聚合物泵。包装袋有足够的含量供一次性使用。标签中包含一个射频识别 (RFID)了解更多信息
Have Questions?
货号数量
43937181 unit
货号 4393718
价格(CNY)
598.00
Each
添加至购物车
数量:
1 unit
价格(CNY)
598.00
Each
添加至购物车
此处提供的调节试剂为与 3500 和 SeqStudio Flex 系列基因分析仪兼容的即用型袋。其设计用于聚合物泵的灌注,在聚合物型改变之间以及仪器关闭过程中洗涤聚合物泵。包装袋有足够的含量供一次性使用。标签中包含一个射频识别 (RFID) 标签。袋顶部用塑料膜密封,在安装到仪器前应先将膜撕掉。
仅供科研使用。不可用于诊断程序。
规格
适用于(设备)3500 系列基因分析仪、SeqStudio Flex 系列基因分析仪
产品规格袋装
产品类型调节试剂
数量1 unit
运输条件湿冰
原始材料DNA
Unit SizeEach
内容与储存
将包装袋储存在 2 至 8°C 下。

常见问题解答 (FAQ)

When I started my run on the Applied Biosystems 3500/3500xL Genetic Analyzer, an error occurred: "Unstable Electrophoresis current detected, check for air bubbles" but I don't see any bubbles. What else can cause this?

Some of the other causes of an Unstable Electrophoresis current detected error message are:

1.Leak on the system
2. Polymer that:
-Has expired
-Has been left on the instrument for more than the recommended time
-Is a mixture of expired polymer and non-expired polymer
3. Running Buffer that was:
-On the instrument longer than 2 weeks
-Not filled to the fill line or evaporated below the fill line
4. An arcing event that was not cleaned afterwards using the water wash wizard
5. Not performing regular maintenance on the instrument
6. Hardware issues

Inspect the system for leaks. If you do not see any leaks on the system, perform the wash the pump chamber and channels wizard using the conditioning pouch and place fresh, non-expired polymer and fresh Anode and Cathode buffer containers. If the problem persists, a service call may be required.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Instruments Support Center.

Do I need to replace the buffer every 14 days if the Applied Biosystems 3500/3500xL Genetic Analyzer is not in use?

If the instrument is not in use, it is not necessary to replace the buffer every 14 days. However, the buffer level needs to remain at the fill line to prevent the capillary or array from drying out. Top off the buffer volume with water if necessary. Replace the anode and cathode buffer containers prior to starting a new run.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Instruments Support Center.