The internally uncoated capillaries are supplied in preassembled sets of 8 capillaries per array, with a built-in frame for easy installation.

Applications of the 50 cm capillaries:
Sequencing fragment analysis and protein quality analysis

For Research Use Only. Not for use in diagnostics procedures.
仅供科研使用。不可用于诊断程序。
规格
毛细管长度50 cm
适用于(应用)蛋白质量分析、片段分析、测序
适用于(设备)3500 基因分析仪
产品规格8 道毛细管阵列
数量1 array
运输条件室温
测试时间125 min。
产品类型3500 基因分析仪 8 道毛细管阵列
Unit SizeEach
内容与储存
在环境温度下储存。有关毛细管阵列储存(安装在仪器上和从仪器上卸载的情况下)的详细信息,请参阅 Applied Biosystems 3500/3500xL 基因分析仪用户指南。

常见问题解答 (FAQ)

How can I set up auto-analysis for data in GeneMapper with the 3500 Data Collection Software version 3.x?

Beginning with 3500 Data Collection Software version 3, the 3500 Data Collection Software does not support automatic secondary analysis. This is also indicated in the software release notes. To perform secondary data analysis, you will need to manually import the .fsa file into the GeneMapper software. For optimal performance, we recommend that you install the GeneMapper Software on a stand-alone computer that is not part of the 3500 Genetic Analyzer system.

Find additional tips, troubleshooting help, and resources within our Fragment Analysis Software Support Center.

I have a 3500 Genetic Analyzer and 14 injections remaining of a 3500 Genetic Analyzer 8-Capillary Array (Cat. No. 4404685). Can I use the capillary array after the expiration date?

We do not recommend using the capillary array after the expiration date.

Clarifications on expiration dates can be found on page 37 of the Applied Biosystems 3500/3500xL Genetic Analyzer User Guide.

Find additional tips, troubleshooting help, and resources within our Next-Generation Sequencing Support Center.

My array is stuck in the pump on my Applied Biosystems 3500/3500xL Genetic Analyzer, even after loosening the knob. What can I do?

If the array is stuck in the pump, even after loosening the knob, squirt around the edge of the array head with distilled water and try to remove it again. If the array will not move, a service call may be required. Do not use excessive force or tools to try and remove it.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Instruments Support Center.

When I started my run on the Applied Biosystems 3500/3500xL Genetic Analyzer, an error occurred: "Unstable Electrophoresis current detected, check for air bubbles" but I don't see any bubbles. What else can cause this?

Some of the other causes of an Unstable Electrophoresis current detected error message are:

1.Leak on the system
2. Polymer that:
-Has expired
-Has been left on the instrument for more than the recommended time
-Is a mixture of expired polymer and non-expired polymer
3. Running Buffer that was:
-On the instrument longer than 2 weeks
-Not filled to the fill line or evaporated below the fill line
4. An arcing event that was not cleaned afterwards using the water wash wizard
5. Not performing regular maintenance on the instrument
6. Hardware issues

Inspect the system for leaks. If you do not see any leaks on the system, perform the wash the pump chamber and channels wizard using the conditioning pouch and place fresh, non-expired polymer and fresh Anode and Cathode buffer containers. If the problem persists, a service call may be required.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Instruments Support Center.

I have an Applied Biosystems 3500/3500xL Genetic Analyzer. When do my consumables expire?

If we do not specify the expiry date or number of uses, the warranty will last for 12 months from the date of shipment.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Instruments Support Center.