阵列卡校准安装试剂盒
Product Image
Applied Biosystems™

阵列卡校准安装试剂盒

本试剂盒用于维护带有 TaqMan™ 阵列卡模块的 ViiA™ 7 系统。它包括使用 FAM™、VIC™了解更多信息
Have Questions?
货号数量
44322501 kit
货号 4432250
价格(CNY)
64,851.00
Each
添加至购物车
数量:
1 kit
价格(CNY)
64,851.00
Each
添加至购物车
本试剂盒用于维护带有 TaqMan™ 阵列卡模块的 ViiA™ 7 系统。它包括使用 FAM™、VIC™ 和 ROX™ 染料进行光谱校准、用于目标区域 (ROI) 和归一化校准以及运行 RNase P 验证卡的试剂。校准和验证操作应至少每六个月运行一次。有关完整的使用说明,请参阅您的仪器维护指南。

您的 ViiA™ 7 系统所用的染料配方将通过更准确地呈现实时荧光定量 PCR 实验中观察到的荧光光谱改善多通路结果。

仅供研究使用。不得用于任何动物或人类的治疗或诊断。
仅供科研使用。不可用于诊断程序。
规格
校准类型光谱校准
适用于(应用)实时 PCR (qPCR)
适用于(设备)ViiA™ 7 系统
数量1 kit
产品类型阵列卡校准安装试剂盒
Unit SizeEach
内容与储存
将试剂储存在 -15°C ± 25°C 下

常见问题解答 (FAQ)

I cannot open the QuantStudio 3/5 Design & Analysis 2 (DA2) software on my Mac. What are the software requirements for Mac?

To open the software download on your Mac, you will need to adjust your security settings using the following instructions from Apple support:

Safely open apps on your mac

Please note, the software should run on Mac OS 10.10 or later.

Find additional tips, troubleshooting help, and resources within our QuantStudio 3 and QuantStudio 5 Real Time PCR Systems Support Center.

How can I get a summary of the calibrations performed on the QuantStudio 6 and 7 Flex Real-Time PCR Systems?

You can get a summary of the calibrations from the QuantStudio Software by going to Instrument > Instrument Console > Manage Instrument > Information > Print Calibration Log.

Find additional tips, troubleshooting help, and resources within our Real-Time PCR and Digital PCR Instruments Support Center.

Do I need to back up my calibration files before I upgrade the software for the ViiA7 Real-Time PCR System?

No, you do not need to back up the calibration files as all of the calibration files are stored on the ViiA7 Real-Time PCR instrument. Once the computer software is properly installed and the computer is connected to the instrument, the calibration information will automatically update in the QuantStudio Software.

Find additional tips, troubleshooting help, and resources within our Real-Time PCR and Digital PCR Instruments Support Center.

Can I still start a run on the ViiA7 Real-Time PCR System if the "Calibration is expired" notification is displayed?

Yes, you can still start a run on the ViiA7 Real-Time PCR System if the expired calibration notice displays. This is just a reminder and you can close it to proceed with the run. Additionally, you can change when you want the notifications to show up and how frequently you want to be notified to calibrate the instrument. To do this, go to the toolbar in the QuantStudio Software and select Instrument > Instrument Console. Select your instrument icon and click on Manager Instrument > Maintenance > Calibration Reminders. From the Calibration reminders Settings page you can update when the calibrations expire, turn expiration reminders on/off, and set when the reminders show up.

Find additional tips, troubleshooting help, and resources within our Real-Time PCR and Digital PCR Instruments Support Center.