Applied Biosystems TaqMan 拷贝数检测试剂盒采用金标准 TaqMan MGB 探针化学试剂,使用 Applied Biosystems 实时了解更多信息
Have Questions?
更改视图buttonViewtableView
货号数量
4442520M (750 reactions), made to order
4442487S (360 reactions), made to order
4442488L (2900 reactions), made to order
货号 4442520
价格(CNY)
8,329.00
Each
立刻订购
数量:
M (750 reactions), made to order
Applied Biosystems TaqMan 拷贝数检测试剂盒采用金标准 TaqMan MGB 探针化学试剂,使用 Applied Biosystems 实时 PCR 仪器和软件来评估基因组 DNA 靶标的拷贝数。TaqMan 拷贝数检测试剂盒在双重 qPCR 反应中与 TaqMan 拷贝数参比检测试剂盒一起运行;拷贝数检测试剂盒检测靶标序列,而参比检测试剂盒已知存在于二倍体基因组的两个拷贝中的序列。然后使用 Applied Biosystems CopyCaller 软件通过相对定量法分析结果。

Custom Plus TaqMan 拷贝数检测是研究无法进行预设计测定的人和小鼠基因组目标区域变异的理想解决方案。这些检测使用免费的 GeneAssist 拷贝数检测工具,采用了开发预设计系列检测所用的相同生物信息学管线。Custom Plus 检测的设计过程包括前期生物信息学序列分析和计算机模拟质量控制。

优点:
灵活性—使用免费的 GeneAssist 拷贝数检测工具生成针对预设计检测系列不包含的人和小鼠靶标的检测方案
特异性—金标准 TaqMan 化学检测及小沟结合物 (MGB) 探针通过使用更短的探针增加检测特异性
重现性—利用我们生物信息学管线开发的稳健检测方案可提供准确、可重现且可靠的结果
易于解读—CopyCaller 软件提供计算的拷贝数和预测的拷贝数,以及置信值和 Z 分数质量指标
快速简便—从构建到获得初步分析结果只需 3–4 小时

预计发货时间
北美 4–6 天,欧洲 6–10 天

所有的检测方案都是我们业界领先的生物信息学管线的产品,这些方案经历了十多年的优化,对制造和检测性能数据加以充分利用。TaqMan 检测已被超过 40,000 篇出版物引用,并且获得超过 350 项专利的支持。

TaqMan 拷贝数参比检测单独出售。

推荐的预混液(单独销售):TaqMan 基因分型预混液

仅供科研使用。不可用于诊断程序。
规格
最大浓度20X
反应次数750 reactions
产品线TaqMan™
数量M (750 reactions), made to order
运输条件Room Temperature
Unit SizeEach
内容与储存
1 tube containing a 20X (S and M sizes) or 60X (L size) mix of pre-formulated assay (1 probe and 2 primers).

Store at -15 to -25°C.

常见问题解答 (FAQ)

How can I reorder a Custom TaqMan Assay that I ordered previously?

To reorder the same Custom TaqMan Assay, you can follow these steps:
- Select an appropriate catalog number, considering the size of the assay that you want. You can find the different catalog numbers for Custom TaqMan Assays on their respective product pages on thermofisher.com.
- Once you have identified the necessary catalog number, go to www.thermofisher.com and click on "Quick Order", in the top right corner of the screen.
- In the following page, paste your selected catalog number in the "Catalog Number" field and paste the assay ID from your previous order in the "Assay ID" field.
- Click "Add to cart" and then in the following screen click "Proceed to check out".

Why do the values for the reference assay dye (VIC dye) vary for my TaqMan Copy Number Assay?

Please review the following possible causes and solutions:
-The reference sequence is not present or contains polymorphisms in the gDNA sample. Use an alternate copy number reference assay.
- There is variability in the amount of gDNA sample added to each reaction. Quantify the gDNA, then adjust the concentration of gDNA as needed.
Note: In general, the calculation of sample-level ΔCt accounts for variability in sample concentration.
- Pipetting was inaccurate. Check the pipette calibration of the pipettes, and pipette at least 5 µL of sample to prepare the reaction mix.

Why is the multicomponent signal for ROX dye not flat for my TaqMan Copy Number Assay?

Most likely incorrect dyes were selected for each target. Check the dyes selected for each target, then reanalyze the data.

Why was there a simultaneous increase in fluorescence from both the passive reference dye (ROX dye) and the reporter dyes for my TaqMan Copy Number Assay?

Most likely the sample evaporated. We recommend that you check the seal of the adhesive film for leaks before running the plate on the real-time PCR instrument.

Why was there a decrease in ROX dye fluorescence (passive reference dye) in my TaqMan Copy Number Assay?

Please review the following possible causes and solutions:
- There was insufficient Master Mix added to the reaction. Please follow accurate pipetting practices.
- Reagents are degraded. Ensure that the kits and reagents have been stored according to the instructions on the packaging and that they have not expired.