我看到KingFisher 24深孔磁头的处理体积为5mL。这是否意味着我可以处理高达5mL的样本?

不是,处理体积指的是24孔深孔板内可以装入的最大体积。这包括结合缓冲液的体积和磁珠体积,所以当扩大反应体积时,一定要确保总体积不超过5mL。

使用KingFisher Flex完成核酸纯化的程序后,磁棒套上仍然有残留的磁珠怎么办?

这种情况有时会发生,但是这不会影响产量,因为样本已经从磁珠上被洗脱下来了。

如果我忘记把KingFisher Flex仪器内的磁棒套插入磁棒套架怎么办?

如果没有向磁棒套架插入磁棒套,则程序不会开始运行。

我的KingFisher Flex磁力棒脏了。我该怎么办?

将软布或纸巾用温和的去垢剂、肥皂水或酒精润湿,然后擦拭磁力棒。

如何替换KingFisher Flex磁棒头?

可以通过运行“维护 (Maintenance)”菜单之下的“更换磁头 (Change magnet)”程序实现,使用“上 (up)”和“下 (down)”按钮进行选择。

KingFisher Flex的磁性会变弱吗?如果是,它们可以使用多久?

The KingFisher Flex磁体是使用非常稳定的材料制成的。磁场不会变弱。不过,极端的机械外力或加热可能会对磁体造成损坏。

注:务必使KingFisher Flex的磁棒头远离其他磁棒或其他磁体。磁体相互接触会引起严重损坏。

我可以在KingFisher Flex 磁珠处理仪上使用其它制造商生产的板子吗?

不行,我们强烈推荐您使用MagMAX板或KingFisher板,因为它们是专为配合KingFisher Flex磁棒套同时使用而设计的,以达到最佳效果。其它制造商生产的板子可能与KingFisher Flex加热模块不兼容。它们还可能引起意料之外的问题,例如可能会由于不同的孔容积和板底高度而引起交叉污染。

KingFisher Flex磁珠处理仪需要配套购买哪些耗材?

带96深孔磁棒头的KingFisher Flex仪器和MagMAX Express-96仪器所需的耗材是相同的。列表如下:

- MagMAX Express-96 深孔板 (货号 4388476)
- MagMAX Express-96 标准板(货号4388475)
- MagMAX Express-96深孔磁棒套(货号4388487)
- MicroAmp 透明封口膜(可选): (货号. 4306311)

对于带24深孔磁棒头的KingFisher Flex仪器,所需耗材列表如下:

- KingFisher Flex 24深孔板:(货号. 95040480)
- KingFisher Flex 24深孔磁棒套: (货号 97002610)

KingFisher Flex 磁珠处理仪有多少种磁棒头选择?

KingFisher Flex 磁珠处理仪有四种磁棒选择:

- 用于MagMAX Express-96深孔板的KingFisher Flex磁棒
- 用于KingFisher 24深孔板的KingFisher Flex磁棒
- 用于KingFisher 96孔板的KingFisher Flex磁棒
- 用于有裙边PCR板的KingFisher Flex磁棒

所有Applied Biosystems MagMAX试剂盒都能用于Thermo Scientific KingFisher Flex 磁珠处理仪吗?

是的,所有MagMAX试剂盒都能用于KingFisher Flex 磁珠处理仪。

Are the KingFisher Plastics for 96 deep-well format (Cat. No 97002534) sterile?

No, that specific catalog number (Cat. No 97002534) is not sterile and purity certified. However, there are other KingFisher plates and tip combs which are sterile and purity certified (see the following: Cat. No. 95040460, 97002820, 95040480 and 97002620).

Find additional tips, troubleshooting help, and resources within our Automated Nucleic Acid Purification Support Center.

What alternative do you offer for KingFisher Deep Well 96 Tip Comb, 100 PCS (Compatible with Flex and Presto) (Cat. No. A48438)?

The alternative is KingFisher Plastics for 96 deep‐well format (Cat. No. 97002534). This product is a 96 tip comb for deep‐well magnets that can be used on the KingFisher Flex and Presto instruments.

Find additional tips, troubleshooting help, and resources within our Automated Nucleic Acid Purification Support Center.

I received the KingFisher Flex/KingFisher Duo Prime tip combs and the retaining clips are bent or broken off. When I try to unbend the retaining clips, they break. Can I still use the tip combs if the retaining clips are bent or broken off?

Do not use the tip combs if the clips are bent or missing. Not having clips can cause the tip comb to drop from the holder during the run, which would cause the run to abort. If you received tip combs with bent or broken retaining clips, you can contact Technical Support at techsupport@thermofisher.com to request a free-of-charge replacement.

Find additional tips, troubleshooting help, and resources within our Automated Nucleic Acid Purification Support Center.

What should I do if the tip comb does not load correctly onto my Thermo Scientific KingFisher system?

If the tip does not load correctly onto a Thermo Scientific KingFisher magnetic head, we recommend that you try manually flexing the tips both lengthwise and widthwise to level the tip comb. A video of flexing the tip comb can be found here: https://www.thermofisher.com/us/en/home/life-science/dna-rna-purification-analysis/automated-purification-extraction/kingfisher-systems/resources.html. The video is under "KingFisher Flex" and titled "How to install the magnet head and check the tip comb." Flexing of the tip comb is demonstrated at timepoint 1:35. If that does not work, try a different box or lot of tips. This applies to all KingFisher instruments.

What is the advantage of using the KingFisher Flex instrument with the 24 Deep-Well magnetic head?

The KingFisher Flex instrument with the 24 Deep-Well magnetic head can process larger volumes for each sample (up to 5mL total volume) compared to the KingFisher Flex instrument with the 96 Deep-Well magnetic head.

Can I use the same tips and plates for the KingFisher Flex instrument with the 96-well magnetic head and the MagMAX Express-96 instrument magnetic head?

Yes, the KingFisher Flex instrument 96-well magnetic head and the MagMAX Express-96 instrument magnetic head can use the same tip combs and KingFisher plates even though the magnetic heads are different.

Do you offer an alternative for the discontinued Pharma MagMAX Express-96 Deep-Well Tip Combs (Cat. No. A31537)?

Yes, we recommend using KingFisher 96 tip combs for deep-well format (Cat. No. 97002534 or 97002570).

Are KingFisher Plastics for 96-deep well format autoclavable?

No, KingFisher Plastics for 96-deep well format are not autoclavable. They won't melt, but they will get deformed.

Can I reuse the Kingfisher Plastics for 96 deep-well format?

The Kingfisher Plastics for 96 deep-well format are recommended for one-time use only, and we don't recommend reusing these products.

Are the KingFisher Plastics for 96 deep-well format CE marked?

The KingFisher Plastics for 96 deep-well format are not CE marked.

Do I have to buy any additional consumables along with the MagMAX kits if I want to use them on the KingFisher Flex instrument?

Yes, depending on the kind of protocol and head that is used, the following consumables will have to be purchased:

- KingFisher 96 deep-well plate (Cat. No. 95040450)
- KingFisher 96 deep-well plate, sterile (Cat. No. 95040460)
- KingFisher 96 tip comb for deep-well (Cat. No. 97002534)
- KingFisher 96 deep-well tip comb and plate, sterile (Cat. No. 97002820)
- KingFisher 96 microplate (200 μL) (Cat. No. 97002540)
- KingFisher 96 tip comb (Cat. No. 97002524)
- MicroAmp Clear Adhesive Film (optional) (Cat. No. 4306311

I see that the KingFisher 24 Deep Well Head has a processing volume of 5 mL. Does that mean that I can process up to 5 mL of my sample?

No, the processing volume refers to the maximum volume that can be held in the 24 Deep Well plates. This incorporates the volume of binding buffers and beads, so when scaling up reaction volumes, it is important to make sure that the total volume does not exceed 5 mL.

My KingFisher Flex run was stopped due to an error message. Can I start the protocol from where it stopped?

No, once a protocol was stopped, it will restart from the beginning. The extraction procedure can either be finished manually by using a magnetic stand, or a modified version of the original KingFisher Flex script would have to be created that starts off at the point where the run was stopped.

I have a 12-plate protocol but there are only 8 slots on the KingFisher Flex instrument turntable. What should I do with the remaining 4 plates?

Set up the protocol on the BindIt software. After the first 8 plates are done, the instrument will pause and instruct you to switch out the plates.

Is it always necessary to use the transport locks on the KingFisher Flex instrument?

The transport locks are only necessary when relocating the instrument.

Are the volumes of reagents in each well critical for a successful KingFisher Flex run?

We strongly recommend that you maintain the specified volumes within the defined limits to avoid spillover and to help ensure the best performance at the most efficient level.

What if the magnetic particles are attached to the tip combs after my KingFisher Flex run?

This happens sometimes but it will not affect the yield because the sample has been released from the particles.

What if the magnetic particles remain in the sample well during a KingFisher Flex run?

If the starting material is too viscous, the magnetic rods will not be able to collect the particles. Dilute the sample and check that the sample is properly homogenized/lysed. Adding low amounts of detergent will improve the collection of magnetic particles as well. Quickly centrifuge the plate to sediment the particles to the bottom of the plate.

What should I do if I forget to insert the tip combs into the tip comb holder for the KingFisher Flex instrument?

The protocol will not start without the tip combs inserted into the tip comb holder.

What should I do if the KingFisher Flex magnet sets are dirty?

Wipe the magnetic rods with a soft cloth or tissue paper soaked in a mild detergent solution, soap solution, or alcohol.

How long does it take the KingFisher Flex heater to warm up from room temperature to 80°C?

The heating block warms up at the rate of about 10°C per minute, so it will take about 6 minutes.

Can I concentrate samples during a KingFisher Flex run?

Both deep-well plates and KingFisher 96 plates can be used during the same run. Therefore, it is possible to start the processing by using larger volumes (in a deep-well plate) and elute the purified sample to a smaller volume (in a KingFisher 96 or MagMAX 96 standard plate).

How can the KingFisher Flex heads be changed?

This can be done by running the Change magnet protocol under the Maintenance menu using the “up” and “down” cursor keys.

Will the KingFisher Flex magnets get weaker? If so, how long can they be used?

The KingFisher Flex magnets are made of material that is very stable. The magnetic field will not get weaker. However, extreme mechanical force or heating can cause damage to the magnets.
Caution: It is very important to keep the KingFisher Flex heads away from each other and from other magnets at all times. Clashing together of the magnets may cause serious damage to the magnets.

How does the KingFisher Flex instrument connect to a computer?

KingFisher Flex uses a USB port, whereas the MagMAX Express- 96 requires a RS-232 port.

I have a custom protocol stored in my KingFisher Flex instrument. Can I edit this protocol and save it to my computer?

No. Protocols cannot be transferred back from the instrument to the computer. Protocols need to be previously stored in the BindIt or KingFisher software to make edits.

Can I edit the manufactured KingFisher Flex protocols?

No, manufactured protocols are locked for editing, but they can be renamed using the ‘save as' function and edited as a new version.

Can I use magnetic beads from other vendors with the KingFisher Flex instrument?

For nucleic acid extraction, we recommend using our MagMAX kits and MagJET kits. Other magnetic beads should work well as long as they are 0.5 to 10 µm in size. However, you will have to write your own KingFisher protocols using the Thermo Scientific BindIt software and validate/optimize the protocol on your own.

Find additional tips, troubleshooting help, and resources within our Automated Nucleic Acid Purification Support Center.

I am working with liquid samples with high volume and want to extract nucleic acid from them using the KingFisher Flex instrument. Can I use more volume than the protocol advises?

Yes, this can be done by sample splitting. Set the protocol so that the magnet will collect samples from multiple wells before continuing to the next step. Be sure not to overload.

Find additional tips, troubleshooting help, and resources within our Automated Nucleic Acid Purification Support Center.

Can I scale up my reaction volume with the KingFisher Flex instrument?

Yes, reaction volumes can be scaled up, and all other volumes (i.e. wash volumes) will have to be increased by the same fold difference. It is generally not recommended to scale up reaction volumes more than 2-fold unless the plate's volume capacity will allow it.

Can I use plates from other manufacturers with the KingFisher Flex Magnetic Particle Processor?

No, we strongly recommend that you use the MagMAX plates or KingFisher plates since they were specifically designed to be used with KingFisher Flex tip combs to attain maximal performance. Plates from other manufacturers may not be compatible with the KingFisher Flex heating blocks. They may also cause unexpected problems, such as cross-contamination, due to the divergent well volume and bottom height of the plate.

What consumables do I need to buy for using the KingFisher Flex Magnetic Particle Processor?

The consumables for the KingFisher Flex instrument with the 96 Deep Well head are listed below:

- KingFisher 96 deep-well plate (Cat. No. 95040450)
- KingFisher 96 deep-well plate, sterile (Cat. No. 95040460)
- KingFisher 96 tip comb for deep-well (Cat. No. 97002534)
- KingFisher 96 deep-well tip comb and plate, sterile (Cat. No. 97002820)
- KingFisher 96 microplate (200 μL) (Cat. No. 97002540)
- KingFisher 96 tip comb (Cat. No. 97002524)
- MicroAmp Clear Adhesive Film (optional) (Cat. No. 4306311

For the KingFisher Flex instrument with the 24 Deep Well head, the following consumables are required:

- KingFisher 24 deep-well plate (Cat. No. 95040470)
- KingFisher 24 deep-well plate, sterile (Cat. No. 95040480)
- KingFisher 24 deep-well tip comb and plate (Cat. No. 97002610)
- KingFisher 24 deep-well tip comb and plate, sterile (Cat. No. 97002620)

How many magnetic head options are available for the KingFisher Flex Magnetic Particle Processor?

There are 4 magnetic head options available for the KingFisher Flex Magnetic Particle Processor:

- KingFisher Flex head for MagMAX Express-96 Deep Well plate
- KingFisher Flex head for KingFisher 24 Deep Well plate
- KingFisher Flex head for KingFisher 96 plate
- KingFisher Flex head for PCR plate, skirted

Are all of the Applied Biosystems MagMAX kits validated for the KingFisher Flex Magnetic Particle Processor?

Yes, all MagMAX kits can be used with the KingFisher Flex Magnetic Particle Processor.