PrestoBlue™ and PrestoBlue™ HS Cell Viability Reagents
Invitrogen17万+抗体限时买二赠一,靶点广,灵活用!
Invitrogen17万+抗体限时买二赠一,靶点广,灵活用!
PrestoBlue™ and PrestoBlue™ HS Cell Viability Reagents
PrestoBlue™ and PrestoBlue™ HS Cell Viability Reagents
PrestoBlue™ and PrestoBlue™ HS Cell Viability Reagents
Invitrogen™

PrestoBlue™ and PrestoBlue™ HS Cell Viability Reagents

Monitor cell proliferation and viability with the resazurin-based PrestoBlue and PrestoBlue HS cell viability reagents. These nontoxic and cell permeable reagents quantitatively measure cell viability in as little as 10 minutes using absorbance- or fluorescence-based microplate readers.
Have Questions?
更改视图buttonViewtableView
货号数量纯度或质量等级
A1326125 mLStandard
A13262100 mLStandard
P5020025 mLHS:
P50201100 mLHS:
货号 A13261
价格(CNY)
3,122.00
Each
添加至购物车
数量:
25 mL
纯度或质量等级:
Standard
价格(CNY)
3,122.00
Each
添加至购物车
PrestoBlue and PrestoBlue HS cell viability reagents are ready-to-use, nontoxic, resazurin-based solutions that function as long term kinetic cell health indicators for fluorescence and absorbance-based microplate assays. These reagents are cell permeable and use the reducing power of living cells to quantitatively measure cell proliferation and viability in as little as 10 minutes. PrestoBlue HS Cell Viability Reagent is an improved version of PrestoBlue reagent, containing highly purified resazurin, which increases sensitivity and reduces background in cell proliferation assays. Add the power of multiplexing with reagents such as the CyQUANT Direct assay to improve your data analysis.

Measuring changes in cell viability is a fundamental method for assessing cell health, determining genotoxicity, and evaluating anti-cancer drugs. Cell health can be monitored by detecting changes in several key indicators, including changes to plasma membrane integrity, DNA synthesis, DNA content, enzyme activity, presence of ATP, and cellular reducing environment.

Features of PrestoBlue and PrestoBlue HS cell viability reagents include:

  • Optimized for reliable detection of mammalian cell viability in only 10 minutes
  • Measure cell viability and proliferation in various cell types, including mammalian (2D and 3D), bacteria, and yeast cells
  • A convenient add-and-read format—no mixing, washing, or cell lysis required
  • Compatible with either fluorescence- or absorbance-based microplate readers
  • Obtain high quality cell proliferation results with a large dynamic range
  • Monitor living cells using real time kinetics and continue culturing them for downstream multiplexed functional assays

Quick 10-minute incubations

Detect mammalian cell viability after just a 10-minute incubation with PrestoBlue and PrestoBlue HS cell viability reagents. These are the only live-cell viability reagents on the market that provide robust data in 10 minutes following reagent addition. All other resazurin-based cell viability reagents require a 1–4 hour incubation with the reagent prior to obtaining data. For best results using 3D samples, it is recommended to incubate the samples for 18–24 hours prior to measurements.

Exceptional sensitivity and accuracy

PrestoBlue and PrestoBlue HS cell viability reagents can detect as few as 12 cells per well in a 384-well plate, depending on the cell type and incubation time. After a 10-minute incubation, PrestoBlue reagent can linearly detect cells down to 98 cells/well with a Z' value greater than 0.5.

Simplified assay format

PrestoBlue and PrestoBlue HS cell viability reagents are addition-only reagents, eliminating the need for cell lysis or solubilization steps required by other assays. This minimizes hands-on time and simplifies the process.

Nontoxic live cell assays

PrestoBlue and PrestoBlue HS cell viability reagents are nontoxic and do not require cell lysis, allowing cells to be incubated for up to 24 hours. Since no lysis is required, the diluted PrestoBlue or PrestoBlue HS solution can be removed from the cells and replaced with complete growth medium for further culturing and use in downstream assays.

PrestoBlue HS reagent—a highly purified version of PrestoBlue reagent with improved performance

All resazurin-based reagents contain a detectable amount of resorufin contamination due to the synthesis and manufacturing process, which contributes to high background fluorescence and reduces performance in cell viability assays. To reduce background and improve performance, an innovative purification process was developed to remove the contaminating resorufin, resulting in the highly pure resazurin used in the PrestoBlue HS Cell Viability Reagent. This highly purified, nontoxic resazurin reduces background fluorescence by >50% and increases signal-to-background by >100% in cell viability microplate assays.

Features of PrestoBlue HS Cell Viability Reagent include:

  • Removal of contaminants from resazurin, resulting in >50% reduction in background fluorescence
  • Increased signal-to-background ratio by >100%, providing a large assay signal window
  • Offers high sensitivity with a linear response, detecting as few as 10 cells per well

Mechanism of PrestoBlue and PrestoBlue HS reagents

Using resazurin-based reagents to monitor changes in cellular reducing environment or metabolic activity is a well-established and reliable method to indicate cell viability or death. PrestoBlue and PrestoBlue HS reagents contain resazurin, which is a nontoxic, cell permeable compound that is blue in color and virtually nonfluorescent. When added to cells, the resazurin-based PrestoBlue and PrestoBlue HS reagents enter cells and are modified by the reducing environment of viable cells. In live cells, resazurin is reduced to resorufin, a compound that is red in color and highly fluorescent. Viable cells continuously convert resazurin to resorufin, increasing the overall fluorescence of the medium surrounding the cells. The conversion of resazurin to resorufin also results in a pronounced color change, allowing cell viability to be detected using either absorbance- or fluorescence-based plate readers.

仅供科研使用。不可用于诊断程序。
规格
细胞类型细菌、真核细胞、真菌细胞
最大浓度10 X
检测方法比色法、荧光法
染料类型刃天青
形式液体
产品规格96 孔板、试管、384 孔板
孵育时间10分钟至2小时
纯度或质量等级Standard
数量25 mL
运输条件经批准可在室温下或置于湿冰上运输
通量兼容高通量应用
颜色蓝色
发射590-615 nm
Excitation Wavelength Range535-560 nm
适用于(应用)活力测定试剂盒
适用于(设备)微孔板读数仪, 分光光度计, HTS 读数仪
产品线PrestoBlue™
产品类型试剂名称
Unit SizeEach
内容与储存
在 2°C 至 8°C 下避光储存。

常见问题解答 (FAQ)

使用PrestoBlue试剂时,平板读数不稳定,这是什么原因所致?

可能的原因之一是您的PrestoBlue试剂中的染料已经沉淀,导致染料浓度变化。这种情况下,应将试剂加热至37°C,然后振荡或漩涡,确保所有的组分都溶于溶液中。另一个原因是移液问题。请确保你的移液器经过校准且吸液前枪头固定牢固。 

我正在使用PrestoBlue试剂。为什么荧光值异常高,以至于超过仪器的线性范围?

请缩短孵育时间和/或减少实验所用的细胞数量。 

我使用PrestoBlue试剂时的荧光值非常低,这是什么原因所致?

我们建议增加细胞在PrestoBlue试剂中的孵育时间,更改仪器的增益或电压设置,检查仪器的过滤器/波长设置。务必在实验设计时安排一个阳性对照(活细胞)组。  

我使用PrestoBlue试剂时观察到高本底荧光值,这是什么原因所致?

试剂遇光可能分解。务必在避光条件下储存PrestoBlue试剂,且不要长时间将试剂直接暴露于光照下。 

我的PrestoBlue染料好像有沉淀,该如何处理?

由于指示剂是一种多组分溶液,我们建议将冷冻的PrestoBlue试剂加热至37°C,振荡或漩涡,确保所有组分都溶于溶液中。