查看更多产品信息 QuantStudio™ 3D Digital PCR Reagent Kit - FAQs (A26360)
14 个常见问题解答
Yes, there is an option to export the Chip data as .xml on the Export tab in AnalysisSuite software v3.0.
The Poisson Plus algorithm corrects for well-to-well load volume variation on a per Chip basis. This becomes important at higher target concentrations.
You should use the Poisson Plus algorithm for projects that contain QuantStudio 3D Chips with target quantities >2,000 copies/µL.
Yes, please use Cat. No. A27322.
At low concentrations, there is no impact on the data. At higher concentrations (>2,000 copies/µL), there is a slight impact. Using the Poisson Plus algorithm will correct for well-to-well volume variation.
No significant difference has been measured.
Yes, internal studies have shown comparable performance between v1 and v2 Chips.
Yes, Firmware v3.0 will work with all versions of Chips and Lids.
No. No matter which version of Chip Lid is being used, v2 Chips require Firmware v3.0 for proper data analysis.
We recommend that you upgrade the firmware to v3.0 in order to read the 2D barcode; otherwise the .eds files will not have a Chip ID.
No, the v2 Master Mix can only be used with the v2 Chips.
No, the current v1 Master Mix can only be used with the older Chips.
Yes, the current Chips will work with the new v2 Chip Lids.
Yes, the current Lids will work on the new v2 Chips.