Why do I have to use ultra pure water for dilution of the Attune NxT Flow Cell Cleaning Solution?

The Attune NxT Flow Cell Cleaning Solution is filtered to remove particulates to avoid scratching optical surfaces, it is important not to introduce particulates from the water, hence, we recommend using ultra pure water for dilution.

Find additional tips, troubleshooting help, and resources within our Cell Analysis Support Center.

What is the pH of the Attune NxT Flow Cell Cleaning Solution?

The Attune NxT Flow Cell Cleaning Solution is at pH 12. This information is posted in the MSDS.

Find additional tips, troubleshooting help, and resources within our Cell Analysis Support Center.

How do I use the Attune NxT Flow Cell Cleaning Solution?

The directions for use are provided on the vial label: Dilute 1:3 in ultra pure water. Use the solution with the SIP sanitize protocol just prior to each shutdown every day.

Find additional tips, troubleshooting help, and resources within our Cell Analysis Support Center.