Search
Search
View additional product information for MagMAX™ Total Nucleic Acid Isolation Kit - FAQs (AM1840)
13 product FAQs found
只有MagMax96用于微阵列总RNA分离试剂盒(货号AM1839)是兼容的,因为它使用Trizol试剂,大部分来自RNAlater试剂的残留的盐在裂解步骤中被去除,并与上层的RNA水相分开。对于其他的MagMax试剂盒,需要进行优化。高盐携带并不是MagMax RNA分离步骤的最佳选择。
不能,出厂实验方案不能编辑,但是您可以使用“另存为”功能对其重命名,然后将它作为一个新的版本来编辑。
对于核酸提取,我们建议使用我们的MagMAX试剂盒和MagJET试剂盒。其它大小在0.5到10 µm间的磁珠应该也可以使用。但是,您需要自己用Thermo Scientific BindIt软件编写您自己的KingFisher实验方案,并自己验证/优化这一实验方案。
可以,可将样品分装。在进行下一步之前,设置程序使得磁头可以从多个孔内收集样本。确保不要过量加样。
可以,反应体积可以扩大,并且所有其它体积(例如洗涤体积)需要以相同倍数扩大。除非板的容积允许,我们强烈建议反应体积扩大不应超过2倍。
是的,所有MagMAX试剂盒都能用于KingFisher Flex 磁珠处理仪。
Only MagMax96 for Microarrays Total RNA Isolation Kit (Cat. No. AM1839) is compatible, since it uses Trizol reagent and most of the salt carry-over from RNAlater reagent is eliminated during the lysis steps and partitioned from the aqueous RNA layer on top. For other MagMax kits, optimization is needed. High salt carry over is not optimal for MagMax RNA isolation steps.
Find additional tips, troubleshooting help, and resources within our RNA Sample Collection, Protection, and Isolation Support Center.
The following components from the MagMAX Total Nucleic Acid Isolaton Kit can be purchased as standalone products and here are the details:
- NA Binding Beads: Can be purchased separately as DNA Binding Beads for MagMAX-96 DNA Multi-Sample Kit, Cat. No. 4489112 (note that these beads are the same as the beads in Cat. No. AM1840, but the fill volume differs)
- Lysis/binding Solution Concentrate: Can be purchased separately as MagMAX Lysis/Binding Solution Concentrate, Cat. No. AM8500
- Wash Solution 1 Concentrate: Can be purchased separately as MagMAX Wash Solution 1 Concentrate, Cat. No. AM8504
- Wash Solution 2 Concentrate: Can be purchased separately as MagMAX Wash Solution 2 Concentrate, Cat. No. AM8640
No, manufactured protocols are locked for editing, but they can be renamed using the save as' function and edited as a new version.
For nucleic acid extraction, we recommend using our MagMAX kits and MagJET kits. Other magnetic beads should work well as long as they are 0.5 to 10 µm in size. However, you will have to write your own KingFisher protocols using the Thermo Scientific BindIt software and validate/optimize the protocol on your own.
Find additional tips, troubleshooting help, and resources within our Automated Nucleic Acid Purification Support Center.
Yes, this can be done by sample splitting. Set the protocol so that the magnet will collect samples from multiple wells before continuing to the next step. Be sure not to overload.
Find additional tips, troubleshooting help, and resources within our Automated Nucleic Acid Purification Support Center.
Yes, reaction volumes can be scaled up, and all other volumes (i.e. wash volumes) will have to be increased by the same fold difference. It is generally not recommended to scale up reaction volumes more than 2-fold unless the plate's volume capacity will allow it.
Yes, all MagMAX kits can be used with the KingFisher Flex Magnetic Particle Processor.