View additional product information for LTQ XL™ Linear Ion Trap Mass Spectrometer - FAQs (IQLAAEGAAVFACZMAIK)
6 product FAQs found
If you notice poor sensitivity or the presence of large background ions, your instrument is in need of cleaning.
Find additional tips, troubleshooting help, and resources within our Mass Spectrometry Support Center.
You may clean the source housing, sweep cone, ion transfer tube, and the RF lens located behind the ITT.
Find additional tips, troubleshooting help, and resources within our Mass Spectrometry Support Center.
The tune file contains four items: Source parameters, ion optic parameters, AGC targets, and scan time settings.
Find additional tips, troubleshooting help, and resources within our Mass Spectrometry Support Center.
See these charts for details.
Find additional tips, troubleshooting help, and resources within our Mass Spectrometry Support Center.
See this table for details.
Find additional tips, troubleshooting help, and resources within our Mass Spectrometry Support Center.
No. However, it is a good idea to calibrate all scan settings at least once a year. In this way, if the zoom scan setting is used in the future, it will be reasonably easy to calibrate at that time.
Find additional tips, troubleshooting help, and resources within our Mass Spectrometry Support Center.