Invitrolon™ PVDF/Filter Paper Sandwiches, 0.45 μm, 8.3 x 7.3 cm (for mini gels) - FAQs

View additional product information for Invitrolon™ PVDF/Filter Paper Sandwiches, 0.45 μm, 8.3 x 7.3 cm (for mini gels) - FAQs (LC2005)

16 product FAQs found

我使用Coomassie Blue对PVDF膜进行染色后,发现背景上有奇怪的旋涡状图案。可能原因是什么?

最可能导致旋涡状图案的原因是膜在染色液中绕圈移动。使用定轨摇床会产生这种效果。为了避免出现这种背景图案,我们建议使用摇臂式摇床或结合摇臂和往复动作的摇床,以确保膜在染色液中均匀晃动。

我将蛋白质转印到PVDF膜上后,转印效率很低。你们有何建议?

以下是可能原因和解决方案:

- 转印前对膜的处理不当:应确保使用甲醇或乙醇等极性有机溶剂润湿膜。
- 凝胶与膜的接触不良:应确保用转膜缓冲液浸透滤纸和海绵垫,小心除去组装膜三明治时的所有气泡。凝胶/膜三明治必须牢固装在两部分印迹模块之间。可尝试多加一个海绵垫或将失去弹性的海绵垫换成新的。
- 过度挤压凝胶:过度挤压的一个明显表现是凝胶过于扁平。在三明治被过度挤压的情况下,应适当移除海绵垫,降低对凝胶和膜施加的过多压力然后关闭转印仪。

你们的PVDF和硝化纤维素膜能否用于Li-COR仪器?

可以,我们的PVDF和硝化纤维素膜均可用于Li-COR仪器。

我正在计划做斑点印迹实验,但我的样品中含有乙腈。你们的PVDF和硝化纤维素膜是否耐受乙腈?

我们的PVDF膜可以耐受乙腈,但硝化纤维素膜不能耐受。

我该如何降低PVDF免疫印迹膜的背景?

PVDF膜需要更为严格的封闭步骤。这可通过将封闭剂的浓度增加2-5倍、延长封闭时间并在37℃封闭来实现。封闭剂与无蛋白质结合的位置结合,可防止背景染色,并封闭与膜结合的蛋白质,以减少其与一抗发生非特异性相互作。封闭剂可使用脱脂奶粉、BSA和酪蛋白。

How can I store, strip, and reuse my western blot?

For nitrocellulose or PVDF membrane following Western blot detection using a chemiluminescent or fluorescent substrate system: Following transfer, air dry the membrane and place in an envelope, preferably on top of a supported surface to keep the membrane flat. The blot can be stored indefinitely at -80 degrees C. When ready to reprobe, prewet the PVDF blot with alcohol for a few seconds, followed by a few rinses with pure water to reduce the alcohol concentration. Then proceed as normal with blocking step.

FOR STRIPPING/REPROBING OF MEMBRANES: Harsh protocol (see NOTE below for modifications)

1) Submerge the membrane in stripping buffer (100 mM BME, 2% SDS, 62.5 mM Tris-HCl, pH 6.7) and incubate at 50 degrees C for 30 min with occasional agitation. If more stringent conditions necessary, incubate at 70 degrees C.

2) Wash 2 x 10 min in TBS-T/PBS-T at room temperature.

3) Block the membrane by immersing in 5% blocking reagent TBS-T or PBS-T for 1 hr at room temperature.

4) Immunodetection

NOTE: Often you don't need such harsh conditions to remove antibodies from their proteins. The stringency of one or several of the variables can be decreased: lower the temperature, decrease the time, less BME, less SDS, etc. An especially mild but still often effective stripping protocol is lower pH incubation. Example: pH 2.0 Tris 50-100 mM, 30-60 min incubation (you may do two incubations if you wish). Then rinse and block as usual. If you do not wish to re-use the membrane immediately after stripping, you can store the membrane in plastic wrap (wet, you do not want it to dry out). Another simple, mild stripping buffer is 0.1 M glycine•HCl (pH 2.5-3.0), incubation 30 min to 2 hrs room temperature or 37 degrees C, depending on the antibody.

Find additional tips, troubleshooting help, and resources within our Protein Assays and Analysis Support Center.

Why must Invitrolon PVDF membranes be dried before staining?

Drying the PVDF membrane reduces the background staining that can occur with wet membranes. A dry PVDF membrane is very hydrophobic and doesn't wet well, but the areas where proteins are bound are more easily saturated and stainable.

This will not work with nitrocellulose membranes, and will only work with PVDF membranes stained for a brief period; staining beyond the recommended time will only increase the background and reduce the detectability.

Find additional tips, troubleshooting help, and resources within our Protein Electrophoresis and Western Blotting Support Center.

How do I destain proteins on a PVDF membrane that were stained with SimplyBlue SafeStain?

After staining with SimplyBlue SafeStain, use deionized water for the less strongly retained protein bands on the PVDF membrane.

Increasing methanol or ethanol concentrations up to 70% should destain any remaining bands. You can leave the membrane in the destain indefinitely.

Find additional tips, troubleshooting help, and resources within our Protein Assays and Analysis Support Center.

Can I use ethanol instead of methanol in the NuPAGE transfer buffer (and in transfer buffers in general)?

We find that including ethanol in the transfer buffer is just as effective as including methanol. You may use either of these alcohols at an equivalent concentration.

Find additional tips, troubleshooting help, and resources within our Protein Electrophoresis and Western Blotting Support Center.

How can I reduce high background on PVDF Western blots?

PVDF membranes require more stringent blocking steps. This can be achieved by increasing the concentration of the blocking reagent 2 to 5 fold, increasing the blocking time, and performing the procedure at 37 degrees C. Blocking agents bind to unoccupied sites to prevent background staining and also to membrane-bound proteins, thus reducing non-specific interactions with the primary antibody. We offer WesternBreeze Immunodetection kits with blocking reagents and Pierce Fast Western Blot kits that have been pre-optimized to give low-background blots. Other examples of blocking agents are nonfat dry milk, BSA, and casein, Starting Block, and SuperBlock.

Find additional tips, troubleshooting help, and resources within our Protein Electrophoresis and Western Blotting Support Center.

What are some causes of poor Western transfer efficiency with a PVDF membrane?

There are three main reasons why this may occur.

1) The membrane may not be properly treated prior to transfer. Make sure that the membrane is pre-wetted with a polar organic solvent such as methanol or ethanol. Failure to do so will prevent proteins from forming hydrophobic interactions with the membrane.

2) There may be poor gel-to-membrane contact. Ensure that the filter paper and blotting pads are well saturated with transfer buffer, taking care to remove any bubbles during the assembly of the membrane sandwich. The gel/membrane sandwich must fit securely in the two halves of the blot module. Try adding another pad or replace any pads that have lost their resiliency with new ones.

3) Overcompression of the gel: a good indication of overcompression is if the gel has been excessively flattened. In the event that the sandwich is overcompressed, remove enough pads so that the blotter can be closed without exerting excess pressure on the gel and membrane.

Find additional tips, troubleshooting help, and resources within our Protein Electrophoresis and Western Blotting Support Center.

I stained my PVDF membrane with Coomassie Blue and am seeing strange swirly patterns in the background. What is the cause for this?

The most probable cause of the swirled artifact is the way the membrane is moved around in the staining solution. Using an orbital shaker can cause this effect. To eliminate these background patterns, we recommend switching to a rocker type shaker or combination of rocker and reciprocal motions to ensure even sloshing of the staining solution over the membrane.

Find additional tips, troubleshooting help, and resources within our Protein Electrophoresis and Western Blotting Support Center.

I performed a western transfer onto a PVDF membrane and the transfer efficiency was very poor. Can you offer some tips?

Here are possible causes and solutions:

- The membrane may not be properly treated prior to transfer: Make sure that the membrane is pre-wetted with a polar organic solvent such as methanol or ethanol.
- There may be poor gel to membrane contact: Ensure that the filter paper and blotting pads are well saturated with transfer buffer, taking care to remove any bubbles during the assembly of the membrane sandwich. The gel/membrane sandwich must fit securely in the two halves of the blot module. Try adding another pad or replace any pads that have lost their resiliency with fresh ones.
- Over-compression of the gel: A good indication of over-compression is if the gel has been excessively flattened. In the event that the sandwich is over-compressed, remove enough pads so that the blotter can be closed without exerting excess pressure on the gel and the membrane.

Find additional tips, troubleshooting help, and resources within our Protein Electrophoresis and Western Blotting Support Center.

Are your PVDF and nitrocellulose membranes compatible with the Li-COR instrument?

Yes, both our PVDF and nitrocellulose membranes are compatible with the Li-COR instrument.

Find additional tips, troubleshooting help, and resources within our Protein Assays and Analysis Support Center.

I am planning on doing a dot blot and my sample contains acetonitrile. Can your PVDF or nitrocellulose membrane withstand acetonitrile?

Our PVDF can tolerate acetonitrile but our nitrocellulose cannot.

Find additional tips, troubleshooting help, and resources within our Protein Assays and Analysis Support Center.

How can I reduce background on my PVDF western blots?

PVDF membranes require more stringent blocking steps. This can be achieved by increasing the concentration of the blocking agent 2-5 fold, increasing the blocking time, and performing the procedure at 37 degrees C. Blocking agents bind to unoccupied sites to prevent background staining and also to membrane-bound proteins, reducing non-specific interactions with the primary antibody. Examples of blocking agents are nonfat dry milk, BSA, and Casein.

Find additional tips, troubleshooting help, and resources within our Protein Assays and Analysis Support Center.