Yes, old spare valves do not have this correction factor and the default value 0 is correct. If an old valve replaces the newer version, the correction factor needs to be set to value 0.
No, you do not have to recalibrate the instrument, because the valve performance is so consistent. However you must apply the valve correction factor to the instrument through the FillIt software.
The orifice (OD) is 100 μm.