3500 基因分析仪 8-毛细管阵列,36 cm
3500 基因分析仪 8-毛细管阵列,36 cm
Applied Biosystems™

3500 基因分析仪 8-毛细管阵列,36 cm

内部未包被毛细管以 8 个毛细管/阵列的预装配组提供,包含内置框架,便于安装。36 cm 毛细管的应用:HID⁄ 法医应用。仅供研究使用。不可用于诊断程序了解更多信息
Have Questions?
货号数量
44046831 array
货号 4404683
价格(CNY)
25,148.00
Each
添加至购物车
数量:
1 array
价格(CNY)
25,148.00
Each
添加至购物车
内部未包被毛细管以 8 个毛细管/阵列的预装配组提供,包含内置框架,便于安装。

36 cm 毛细管的应用:
HID⁄ 法医应用。

仅供研究使用。不可用于诊断程序。
仅供科研使用。不可用于诊断程序。
规格
毛细管长度36 cm
描述3500 基因分析仪 8-毛细管阵列,36 cm 毛细管,内部未包被毛细管,用于与 3500 基因分析仪配合使用,储存温度为 15 至 30°C
适用于(应用)Human Identification,Sequencing,Fragment Analysis
适用于(设备)3500 Genetic Analyzer
长度(公制)36 cm(毛细管)
产品类型基因分析仪 8-毛细管阵列
数量1 array
运输条件室温
Unit SizeEach
内容与储存
该阵列在环境温度下储存。有关毛细管阵列储存(安装在仪器上和从仪器上卸载的情况下)的详细信息,请参阅 Applied Biosystems 3500/3500xL 基因分析仪用户指南。

常见问题解答 (FAQ)

How can I set up auto-analysis for data in GeneMapper with the 3500 Data Collection Software version 3.x?

Beginning with 3500 Data Collection Software version 3, the 3500 Data Collection Software does not support automatic secondary analysis. This is also indicated in the software release notes. To perform secondary data analysis, you will need to manually import the .fsa file into the GeneMapper software. For optimal performance, we recommend that you install the GeneMapper Software on a stand-alone computer that is not part of the 3500 Genetic Analyzer system.

Find additional tips, troubleshooting help, and resources within our Fragment Analysis Software Support Center.

My array is stuck in the pump on my Applied Biosystems 3500/3500xL Genetic Analyzer, even after loosening the knob. What can I do?

If the array is stuck in the pump, even after loosening the knob, squirt around the edge of the array head with distilled water and try to remove it again. If the array will not move, a service call may be required. Do not use excessive force or tools to try and remove it.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Instruments Support Center.

When I started my run on the Applied Biosystems 3500/3500xL Genetic Analyzer, an error occurred: "Unstable Electrophoresis current detected, check for air bubbles" but I don't see any bubbles. What else can cause this?

Some of the other causes of an Unstable Electrophoresis current detected error message are:

1.Leak on the system
2. Polymer that:
-Has expired
-Has been left on the instrument for more than the recommended time
-Is a mixture of expired polymer and non-expired polymer
3. Running Buffer that was:
-On the instrument longer than 2 weeks
-Not filled to the fill line or evaporated below the fill line
4. An arcing event that was not cleaned afterwards using the water wash wizard
5. Not performing regular maintenance on the instrument
6. Hardware issues

Inspect the system for leaks. If you do not see any leaks on the system, perform the wash the pump chamber and channels wizard using the conditioning pouch and place fresh, non-expired polymer and fresh Anode and Cathode buffer containers. If the problem persists, a service call may be required.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Instruments Support Center.

I have an Applied Biosystems 3500/3500xL Genetic Analyzer. When do my consumables expire?

If we do not specify the expiry date or number of uses, the warranty will last for 12 months from the date of shipment.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Instruments Support Center.

How often should I change the array on my Applied Biosystems 3500/3500xL Genetic Analyzer?

The Applied Biosystems 3500/3500xl Genetic Analyzer capillary arrays are warrantied for a specific number of runs or until the expiration date printed on the array package, whichever comes first. If you notice that you are starting to get reduced data quality (e.g., shorter reads, trailing peaks in all 4 colors) even with the Sequencing Standard, these may be signs that the array requires changing.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Instruments Support Center.