This 8-tube retainer and base set is for use with standard 8-tube strips on the 3500 series Genetic Analyzers.

For Research Use Only. Not for use in diagnostics procedures.
仅供科研使用。不可用于诊断程序。
规格
描述固定器和底座套装(标准),8 孔联管,0.2 mL,适用于 3500/3500XL 基因分析仪,室温,标准反应速度
适用于(应用)测序
适用于(设备)3500 基因分析仪、3500xL 基因分析仪
产品类型底座套装
数量4 套
样品体积0.2 mL
运输条件室温
测试时间标准
管类型联排管
产品规格8 孔联排管
Unit SizeEach
内容与储存
该产品包括 4 套固定器和底座。

常见问题解答 (FAQ)

When I started my run on the Applied Biosystems 3500/3500xL Genetic Analyzer, an error occurred: "Unstable Electrophoresis current detected, check for air bubbles" but I don't see any bubbles. What else can cause this?

Some of the other causes of an Unstable Electrophoresis current detected error message are:

1.Leak on the system
2. Polymer that:
-Has expired
-Has been left on the instrument for more than the recommended time
-Is a mixture of expired polymer and non-expired polymer
3. Running Buffer that was:
-On the instrument longer than 2 weeks
-Not filled to the fill line or evaporated below the fill line
4. An arcing event that was not cleaned afterwards using the water wash wizard
5. Not performing regular maintenance on the instrument
6. Hardware issues

Inspect the system for leaks. If you do not see any leaks on the system, perform the wash the pump chamber and channels wizard using the conditioning pouch and place fresh, non-expired polymer and fresh Anode and Cathode buffer containers. If the problem persists, a service call may be required.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Instruments Support Center.

Do I need to replace the buffer every 14 days if the Applied Biosystems 3500/3500xL Genetic Analyzer is not in use?

If the instrument is not in use, it is not necessary to replace the buffer every 14 days. However, the buffer level needs to remain at the fill line to prevent the capillary or array from drying out. Top off the buffer volume with water if necessary. Replace the anode and cathode buffer containers prior to starting a new run.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Instruments Support Center.