用于 3500xL 基因分析仪的&固定器和底座套件(标准),384 孔
Product Image
Applied Biosystems™

用于 3500xL 基因分析仪的&固定器和底座套件(标准),384 孔

该 384 孔板固定器和座底标准套件与 3500xL 基因分析仪配合使用。仅供研究使用。不适用于诊断用途了解更多信息
Have Questions?
货号数量
44102354 套
货号 4410235
价格(CNY)
3,984.00
Each
添加至购物车
数量:
4 套
价格(CNY)
3,984.00
Each
添加至购物车
该 384 孔板固定器和座底标准套件与 3500xL 基因分析仪配合使用。

仅供研究使用。不适用于诊断用途。
仅供科研使用。不可用于诊断程序。
规格
适用于(应用)测序
适用于(设备)3500xL 基因分析仪
产品类型底座套装
数量4 套
运输条件室温
产品规格384 孔板
Unit SizeEach
内容与储存
该产品包括 4 套固定器和底座。

常见问题解答 (FAQ)

When I started my run on the Applied Biosystems 3500/3500xL Genetic Analyzer, an error occurred: "Unstable Electrophoresis current detected, check for air bubbles" but I don't see any bubbles. What else can cause this?

Some of the other causes of an Unstable Electrophoresis current detected error message are:

1.Leak on the system
2. Polymer that:
-Has expired
-Has been left on the instrument for more than the recommended time
-Is a mixture of expired polymer and non-expired polymer
3. Running Buffer that was:
-On the instrument longer than 2 weeks
-Not filled to the fill line or evaporated below the fill line
4. An arcing event that was not cleaned afterwards using the water wash wizard
5. Not performing regular maintenance on the instrument
6. Hardware issues

Inspect the system for leaks. If you do not see any leaks on the system, perform the wash the pump chamber and channels wizard using the conditioning pouch and place fresh, non-expired polymer and fresh Anode and Cathode buffer containers. If the problem persists, a service call may be required.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Instruments Support Center.

Do I need to replace the buffer every 14 days if the Applied Biosystems 3500/3500xL Genetic Analyzer is not in use?

If the instrument is not in use, it is not necessary to replace the buffer every 14 days. However, the buffer level needs to remain at the fill line to prevent the capillary or array from drying out. Top off the buffer volume with water if necessary. Replace the anode and cathode buffer containers prior to starting a new run.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Instruments Support Center.