适用于 96 孔板的隔垫,用于 3500/SeqStudio™ Flex
Product Image
Applied Biosystems™

适用于 96 孔板的隔垫,用于 3500/SeqStudio™ Flex

这些适用于 96 孔板的隔垫经过设计和验证,可与 Applied Biosystems 3500 和 SeqStudio Flex 系列基因分析仪配合使用了解更多信息
Have Questions?
货号数量
441261420 each
货号 4412614
价格(CNY)
7,099.00
Each
添加至购物车
数量:
20 each
价格(CNY)
7,099.00
Each
添加至购物车
这些适用于 96 孔板的隔垫经过设计和验证,可与 Applied Biosystems 3500 和 SeqStudio Flex 系列基因分析仪配合使用。
仅供科研使用。不可用于诊断程序。
规格
适用于(设备)3500 基因分析仪、3500xL 基因分析仪
产品规格96 孔板
产品类型隔垫板
数量20 each
运输条件室温
Unit SizeEach
内容与储存
一袋 20 个隔垫。在室温下储存。

常见问题解答 (FAQ)

How often can the plate septa on the 3130/xl, 3730/xl, 3500/xl, and SeqStudio instruments be used?

We officially recommend that the plate septa be used one time only. However, we are aware that some of our customers will rinse the septa with water and re-use them. In those instances, the septa can become sticky and gummy over time and eventually start sticking to the array. In the best case scenario, the array will lift the plate off of the autosampler just enough to break the connection with the plate sensor and abort the run. The Event Log might show that the run aborted due to the user aborting it, so the run will appear to have completed or stopped with no errors and no data. The worst case scenario is that the plate remains on the array as the autosampler starts to move, bending the electrodes and breaking the array. We recommend replacing the plate septa regularly (replacing after each run is recommended). Do not use bleach, detergents, or solvents to clean the septa, it will cause them to dry out faster and also risks contaminating the system.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Software Support Center.

When I started my run on the Applied Biosystems 3500/3500xL Genetic Analyzer, an error occurred: "Unstable Electrophoresis current detected, check for air bubbles" but I don't see any bubbles. What else can cause this?

Some of the other causes of an Unstable Electrophoresis current detected error message are:

1.Leak on the system
2. Polymer that:
-Has expired
-Has been left on the instrument for more than the recommended time
-Is a mixture of expired polymer and non-expired polymer
3. Running Buffer that was:
-On the instrument longer than 2 weeks
-Not filled to the fill line or evaporated below the fill line
4. An arcing event that was not cleaned afterwards using the water wash wizard
5. Not performing regular maintenance on the instrument
6. Hardware issues

Inspect the system for leaks. If you do not see any leaks on the system, perform the wash the pump chamber and channels wizard using the conditioning pouch and place fresh, non-expired polymer and fresh Anode and Cathode buffer containers. If the problem persists, a service call may be required.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Instruments Support Center.

Do I need to replace the buffer every 14 days if the Applied Biosystems 3500/3500xL Genetic Analyzer is not in use?

If the instrument is not in use, it is not necessary to replace the buffer every 14 days. However, the buffer level needs to remain at the fill line to prevent the capillary or array from drying out. Top off the buffer volume with water if necessary. Replace the anode and cathode buffer containers prior to starting a new run.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Instruments Support Center.