ZOOM™ IPGRunner™ 盒
ZOOM™ IPGRunner™ 盒
ZOOM™ IPGRunner™ 盒
ZOOM™ IPGRunner™ 盒
ZOOM™ IPGRunner™ 盒
Invitrogen™

ZOOM™ IPGRunner™ 盒

ZOOM IPGRunner 盒与 ZOOM IPGRunner 小型槽配合使用。其中包含六个可容纳7.0-cm ZOOM 胶条的封闭通道,作为胶条的水化和运行室。每个封闭通道内都含有用于水化了解更多信息
Have Questions?
货号数量
ZM000310 cassette(s)
货号 ZM0003
价格(CNY)
6,488.00
10 cassette(s)
添加至购物车
数量:
10 cassette(s)
价格(CNY)
6,488.00
10 cassette(s)
添加至购物车
ZOOM IPGRunner 盒与 ZOOM IPGRunner 小型槽配合使用。其中包含六个可容纳7.0-cm ZOOM 胶条的封闭通道,作为胶条的水化和运行室。每个封闭通道内都含有用于水化 ZOOM 胶条的水化缓冲液和样品。有两个上样设备与 ZOOM IPGRunner 盒的上端(阴极)和下端(阳极)相连。上样设备中的上样孔是专为样品应用而设计的,可防止任何样品在上样和水化过程中溢出到相邻通道中。
仅供科研使用。不可用于诊断程序。
规格
数量10 cassette(s)
运输条件室温
适用于(设备)ZOOM™ IPGRunner™ 系统
产品线IPGRunner™、ZOOM™
Unit Size10 cassette(s)
内容与储存
• Zoom IPGRunner 盒 (10)
• 密封带 (22)
• 电极芯(22)

常见问题解答 (FAQ)

我刚设置好用于IEF的ZOOM IPGRunner系统,但没有任何电流通过系统。问题出在哪里?

一个可能的原因是电极间连接不良或电路未闭合。应确保在Electrode Wicks中加入600 μL去离子水,并且凝胶暴露于凝胶盒的阳极和阴极端。检查电源。确保关闭电源的“负载检测”功能,使电源能够在低电流下运行。

ZOOM IPG Runner Mini-Cell包含哪些组件?

以下是ZOOM IPG Runner Mini-Cell的组件:

•ZOOM IPGRunner 核心
•ZOOM IPGRunner盖子
•凝胶张力楔
•缓冲液挡板
•Mini-Cell槽

What may cause streaking on the 2nd dimension gel after IEF?

There are several reasons why streaking may occur.

(1) Sample is not completely solubilized prior to application.

(2) Sample is poorly soluble in rehydration solution.

(3) Non-protein impurities in the sample can interfere with IEF, causing horizontal streaking in the final 2-D result, particularly toward the acidic side of the gel.

(4) Ionic impurities are present in sample.

(5) Ionic detergent is present in sample.

(6) Sample load is too high.

(7) Underfocusing. Focusing time was not long enough to achieve steady state focusing.

(8) Overfocusing. Extended focusing times (over 100,000 Vh) may result in electroendosmotic water and protein movement, which can produce horizontal smearing.

What should be done?

(1) Be sure that the sample is completely and stably solubilized. Note: Repeated precipitation-resolubilization cycles produce or increase horizontal streaking.

(2) Increase the concentration of the solubilizing components in the rehydration solution.

(3) Modify sample preparation to limit these contaminants or dialyze protein.

(4) Reduce salt concentration to below 10 mM by dilution or desalt the sample by dialysis. Precipitation with TCA and acetone and subsequent resuspension is another effective desalting technique that removes lipids, nucleotides and other small molecules.
Note: Specific and non-specific losses of proteins can occur with dialysis, gel chromatography, and precipitation/resuspension of samples. If the sample preparation cannot be modified, the effect of ionic impurities can be reduced by modifying the IEF protocol. Limit the voltage to 100-150 V for 2 hours, then resume a normal voltage step program. This pre-step allows the ions in the sample to move to the ends of the IPG strip.

(5) If the ionic detergent SDS is used in sample preparation, the final concentration must not exceed 0.25% after dilution into the rehydration solution. Additionally, the concentration of the non-ionic detergent present must be at least 8 times higher than the concentration of any ionic detergent to ensure complete removal of SDS from the proteins.

(6) Extend focusing time. Load less sample.

(7) Prolong focusing time.

(8) Reduce focusing time.

Find additional tips, troubleshooting help, and resources within our Protein Assays and Analysis Support Center.

I just set up the ZOOM IPGRunner for IEF but don't see any current running through the system. What is the issue?

A possible reason is poor contact between electrodes or incomplete circuit. Make sure that you have added 600 µL deionized water to the Electrode Wicks and the gel is exposed at the anodic and cathodic ends of the cassette. Check the power supply. Be sure to set the “Load Check” to off to enable the power supply to operate at low current.

Find additional tips, troubleshooting help, and resources within our Protein Electrophoresis and Western Blotting Support Center.

What are the components of the ZOOM IPGRunner Mini-Cell?

Here are the components of the ZOOM IPGRunner Mini-Cell:

ZOOM IPGRunner Core
ZOOM IPGRunner Lid
Gel Tension Wedge
Buffer Dam
Mini-Cell Chamber

Find additional tips, troubleshooting help, and resources within our Protein Electrophoresis and Western Blotting Support Center.