NanoDrop™ Reagents
NanoDrop™ Reagents
Thermo Scientific™

NanoDrop™ Reagents

Certified standards for verifying instrument performance to factory specifications are available separately.
Have Questions?
更改视图buttonViewtableView
货号类型
CHEM-CF-1水性重铬酸钾 (K2Cr207) 溶液
CHEM-CF-8校准检查套件
CHEM-PR1-KITNanoDrop PR-1 调节套件
CHEM-PV-1PV-1 Performance Verification
CHEM-PV-8PV-8 Performance Verification
货号 CHEM-CF-1
价格(CNY)
-
申请报价
类型:
水性重铬酸钾 (K2Cr207) 溶液
Certified standards for verifying instrument performance to factory specifications are available separately.

CHEM-PV-1 - PV-1 Performance Verification Solution

For use with NanoDrop Ultra Spectrophotometers and Fluorometers, NanoDrop One/OneC and NanoDrop Lite Plus
Solution with certified absorbance values.
Use to create a record of correct photometric performance.
Intended for single use only.
 

CHEM-PV-8 - PV-8 Performance Verification Kit

For use with NanoDrop Eight
Solution with certified absorbance values.
Use to create a record of correct photometric performance.
Intended for single use only.
 

CHEM-CF-1 - CF-1 Calibration Fluid

For use with NanoDrop 1000, NanoDrop 2000/2000c, NanoDrop Lite
Solution with certified absorbance values.
Use to create a record of correct photometric performance.
Intended for single use only.
 

CHEM-CF-8 - CF-8 Calibration Check Kit

For use with NanoDrop 8000
Solution with certified absorbance values.
Use to create a record of correct photometric performance.
Intended for single use only.
 

CHEM-PR1-KIT - PR-1 Reconditioning Kit

For use with all NanoDrop models
This kit is useful for reconditioning the sample measurement pedestals to a hydrophobic state if they become “unconditioned” (see user manual for more information). The PR-1 kit consists of a small screw cap vial of specially formulated compound and a supply of 25 convenient applicators.

规格
适用于(设备)NanoDrop 8000、NanoDrop Lite、NanoDrop 2000、NanoDrop 2000c 分光光度计
包装0.5 mL 弹簧盖玻璃安瓿瓶
类型水性重铬酸钾 (K2Cr207) 溶液
Unit SizeEach

常见问题解答 (FAQ)

How do I check the accuracy of my NanoDrop 2000/2000c Spectrophotometer?

Our NanoDrop CF-1 Calibration Fluid should be used with the Calibration Check module of the instrument software. CF-1 is prepared from the NIST potassium dichromate standard SRM935 in acidified reagent grade water (Certificate of Analysis available from NanoDrop products).

How often do I need to check the calibration of my NanoDrop Lite Spectrophotometer?

We recommend confirming that your NanoDrop Lite Spectrophotometer is working within specifications every 6 months. The NanoDrop CF-1 Calibration Fluid is required to run the calibration check procedure. CF-1 is available from Thermo Fisher website (Cat. No. CHEM-CF-1)

How do I check the accuracy of my NanoDrop Lite Spectrophotometer?

The accuracy of your NanoDrop Lite Spectrophotometer can be checked by using the NanoDrop CF-1 Calibration Fluid to perform a calibration check. CF-1 is prepared from the NIST potassium dichromate standard SRM935 in acidified reagent grade water.

My NanoDrop 8000 Spectrophotometer failed calibration check. What should I do?

Please review the following steps to take when/if your NanoDrop 8000 Spectrophotometer fails a calibration check:
- Clean the pedestal with the NanoDrop PR-1 Reconditioning Kit
- Repeat calibration check using a FRESH vial of NanoDrop CF-1 Calibration Fluid and FRESH blanking solution
- If instrument continues to be out of specification, Contact Technical Support

How do I check the calibration of my NanoDrop 8000 Spectrophotometer? Do I need to calibrate it?

You can easily check the calibration of your NanoDrop 8000 Spectrophotometer by using the NanoDrop CF-8 Calibration Check Kit, which contains two CF-1 Ampules and two 8-well PCR tube strips. The CF-8 kit is available from the Thermo Fisher website (Cat. No. CHEM-CF-8).
If re-calibration is ever needed, it is performed by a trained service technician. Due to the precise nature of the NanoDrop instrument design, these spectrophotometers rarely require re-calibration.